FDA Medical Device Recalls (openFDA)
Datascope Corporation: Custom Tubing Pack with Bioline Coating UDI code: 00607567208230 Part Number: 709000058 The HLM Tubing Sets with Bioline Coating are for single use only. They may be sold sterile, non-sterile, and bulk packed. Custom tubing sets that are sold sterile are not to be re-sterilized by the user. In open heart surgery the HLM Tubing Sets with Bioline Coating are used in combination with the heart-lung machine for the oxygenation of blood and removal of carbon dioxide. The main purpose of the HLM Tubing Sets with Bioline Coating is to connect the patient to the heart-lung machine and its components. The HLM Tubing Sets with Bioline Coating are therefore a component in the extracorporeal perfusion circulation system. The utilization period of the use of the tubing sets is restricted to six hours. The Bioline Coating improves the physical surface properties of products for the extracorporeal circulation system.
- Recall number
- Z-1769-2017
- Recalling firm
- Datascope Corporation
- Product
- Custom Tubing Pack with Bioline Coating UDI code: 00607567208230 Part Number: 709000058 The HLM Tubing Sets with Bioline Coating are for single use only. They may be sold sterile, non-sterile, and bulk packed. Custom tubing sets that are sold sterile are not to be re-sterilized by the user. In open heart surgery the HLM Tubing Sets with Bioline Coating are used in combination with the heart-lung machine for the oxygenation of blood and removal of carbon dioxide. The main purpose of the HLM Tubing Sets with Bioline Coating is to connect the patient to the heart-lung machine and its components. The HLM Tubing Sets with Bioline Coating are therefore a component in the extracorporeal perfusion circulation system. The utilization period of the use of the tubing sets is restricted to six hours. The Bioline Coating improves the physical surface properties of products for the extracorporeal circulation system.
- Status
- Terminated
- Initiated
- 2017-02-16
- Posted
- 2017-03-23
- Terminated
- 2019-03-21
- Reason
- A complaint was received on January 1, 2017 stating the pigtail on line 2 of the oxygenator was assembled to the port where the yellow cap and the clear cap remained on the center port.
- Root cause
- Device Design
- Action
- Maquet Inc. was notified by their customer on January 1, 2017. This was a reverse notification. The one impacted customer said they will not return the product since they can easily correct the pigtail connection before use. For further questions regarding this recall, please call (973) 709-7442.
- Quantity
- 21 kits
- Distribution
- US Distribution to the state of : CA.
- Lot, serial or model codes
- Lot number 3000033848
- 510(k) numbers
- ["K080592"]
- PMA numbers
- not on file
- Product code
- DWE
- Firm FEI
- 1000222374
- Firm city
- Fairfield
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28