Skip to the content

FDA Medical Device Recalls (openFDA)

Datascope Corporation: Custom Tubing Pack with Bioline Coating UDI code: 00607567208230 Part Number: 709000058 The HLM Tubing Sets with Bioline Coating are for single use only. They may be sold sterile, non-sterile, and bulk packed. Custom tubing sets that are sold sterile are not to be re-sterilized by the user. In open heart surgery the HLM Tubing Sets with Bioline Coating are used in combination with the heart-lung machine for the oxygenation of blood and removal of carbon dioxide. The main purpose of the HLM Tubing Sets with Bioline Coating is to connect the patient to the heart-lung machine and its components. The HLM Tubing Sets with Bioline Coating are therefore a component in the extracorporeal perfusion circulation system. The utilization period of the use of the tubing sets is restricted to six hours. The Bioline Coating improves the physical surface properties of products for the extracorporeal circulation system.

Recall number
Z-1769-2017
Recalling firm
Datascope Corporation
Product
Custom Tubing Pack with Bioline Coating UDI code: 00607567208230 Part Number: 709000058 The HLM Tubing Sets with Bioline Coating are for single use only. They may be sold sterile, non-sterile, and bulk packed. Custom tubing sets that are sold sterile are not to be re-sterilized by the user. In open heart surgery the HLM Tubing Sets with Bioline Coating are used in combination with the heart-lung machine for the oxygenation of blood and removal of carbon dioxide. The main purpose of the HLM Tubing Sets with Bioline Coating is to connect the patient to the heart-lung machine and its components. The HLM Tubing Sets with Bioline Coating are therefore a component in the extracorporeal perfusion circulation system. The utilization period of the use of the tubing sets is restricted to six hours. The Bioline Coating improves the physical surface properties of products for the extracorporeal circulation system.
Status
Terminated
Initiated
2017-02-16
Posted
2017-03-23
Terminated
2019-03-21
Reason
A complaint was received on January 1, 2017 stating the pigtail on line 2 of the oxygenator was assembled to the port where the yellow cap and the clear cap remained on the center port.
Root cause
Device Design
Action
Maquet Inc. was notified by their customer on January 1, 2017. This was a reverse notification. The one impacted customer said they will not return the product since they can easily correct the pigtail connection before use. For further questions regarding this recall, please call (973) 709-7442.
Quantity
21 kits
Distribution
US Distribution to the state of : CA.
Lot, serial or model codes
Lot number 3000033848
510(k) numbers
["K080592"]
PMA numbers
not on file
Product code
DWE
Firm FEI
1000222374
Firm city
Fairfield
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register