Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: System AXIOM Aristos FX Plus - A dedicated x-ray system with a flat panel detector which allows the acquisition of x-ray exposures without the use of conventional film/screen systems; and allows radiographic exposures of the whole body including skull, spinal column, chest, abdomen, and extremities. It is not designed for mammography use. Radiographic exposures may be taken with the patient in the sitting, standing, or supine positions.

Recall number
Z-1765-2020
Recalling firm
Siemens Medical Solutions USA, Inc
Product
System AXIOM Aristos FX Plus - A dedicated x-ray system with a flat panel detector which allows the acquisition of x-ray exposures without the use of conventional film/screen systems; and allows radiographic exposures of the whole body including skull, spinal column, chest, abdomen, and extremities. It is not designed for mammography use. Radiographic exposures may be taken with the patient in the sitting, standing, or supine positions.
Status
Terminated
Initiated
2020-04-06
Posted
not on file
Terminated
2020-10-27
Reason
It was determined that in rare cases persons can receive low dose of unintended radiation. The system performed exposures without the exposure button being pressed. Exposure will be released for less than 500ms, therefore, the overall dose is estimated at 0.01mSv or less than 500 mGy in total.
Root cause
Radiation Control for Health and Safety Act
Action
Only one user in the United States is affected by this corrective action. The Customer Safety Advisory Notice was sent to the affected customer. Any additional communication regarding this corrective action will include the following statement: The manufacturer will, without charge, remedy the defect or bring the product into compliance with each applicable Federal standard in accordance with a plan to be approved by the Secretary of Health and Human Services, the details of which will be included
Quantity
399 (wordwide)
Distribution
Worldwide Distribution
Lot, serial or model codes
serail number 1192
510(k) numbers
["K061054"]
PMA numbers
not on file
Product code
MQB
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register