Teleflex Medical: Sheridan/CF Novaplus 7.0 mm, Product Code V5-10114 Product Usage: Tracheal tube/airway management

Recall number
Z-1765-2019
Recalling firm
Teleflex Medical
Product
Sheridan/CF Novaplus 7.0 mm, Product Code V5-10114 Product Usage: Tracheal tube/airway management
Status
Terminated
Initiated
2019-05-24
Posted
2019-06-19
Terminated
2022-02-15
Reason
This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.
Root cause
Process control
Action
Teleflex sent an Urgent Medical Device Recall letter dated May 24, 2019 to customers. The letter identified the affected product, problem and actions to be taken. The letter directed customers to discontinue use, quarantine the product, and contact Teleflex for return of the product. Distributors were directed to notify their customers and request the customer return the recalled products to them for consolidation and subsequent return to Teleflex.. For questions contact your local sales representative or Customer Service at 1-866-396-2111
Quantity
77136 units
Distribution
Worldwide - US Nationwide Distribution
Lot, serial or model codes
Lot Numbers: 73K1600156 73K1600735 73L1600501 73A1700120 73B1700321 73C1700425 73D1700630 73E1700504 73G1700146 73H1700053 73J1700307 73K1700085 73B1800223 73B1800371 73C1800372 73D1800655 73E1800579 73G1800627 73H1800203 73H1800429 73H1800758 73J1800118 73J1800696
510(k) numbers
["K822082"]
PMA numbers
not on file
Product code
BTR
Firm FEI
3005747797
Firm city
Morrisville
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register