Kyphon Inc: X-STOP Interspinous Process Decompression (lPD) System, (With Physician's Guide, P/N 16000789-01 (Ti device), P/N 16001013-01 (PEEK device); Catalog numbers for X-STOP Titanium Implant: 1-2206 US 6MM X-STOP, 1-2208 US 8MM X-STOP, 1-2210 US 10MM X-STOP, 1-2212 US 12MM X-STOP, 1-2214 US 14MM X-STOP. Catalog numbers include for X-STOP PEEK Implant: 1-3206 US 6mm X STOPpk, 1-3208 US 8mm X STOPpk, 1-3210 US 10mm X STOPpk, 1-3212 US 12mm X STOPpk, 1-3214 US 14mm X STOPpk, 1-3216 US 16mm X STOPpk. Product is manufactured and distributed by Medtronic Spine LLC (formerly Kyphon Inc.); Spinal implants.

Recall number
Z-1765-2008
Recalling firm
Kyphon Inc
Product
X-STOP Interspinous Process Decompression (lPD) System, (With Physician's Guide, P/N 16000789-01 (Ti device), P/N 16001013-01 (PEEK device); Catalog numbers for X-STOP Titanium Implant: 1-2206 US 6MM X-STOP, 1-2208 US 8MM X-STOP, 1-2210 US 10MM X-STOP, 1-2212 US 12MM X-STOP, 1-2214 US 14MM X-STOP. Catalog numbers include for X-STOP PEEK Implant: 1-3206 US 6mm X STOPpk, 1-3208 US 8mm X STOPpk, 1-3210 US 10mm X STOPpk, 1-3212 US 12mm X STOPpk, 1-3214 US 14mm X STOPpk, 1-3216 US 16mm X STOPpk. Product is manufactured and distributed by Medtronic Spine LLC (formerly Kyphon Inc.); Spinal implants.
Status
Terminated
Initiated
2008-01-30
Posted
2008-09-29
Terminated
2009-07-14
Reason
Physician instructions revised due to product breakage: During the implantation of the X-STOP IPD Device, the product has the potential to cause damage to and/or breakage of the X-STOP IPD Device's universal wing assembly. This includes the potential for breakage of the wing screw.
Root cause
Labeling design
Action
On January 30, Kyphon sent a letter to all consignees, and a teleconference was held with sales management informing them of the recall and the need to contact consignees. The product labeling including the Physician Training and Physician's Guide have been updated to reflect the potential risk of breakage.
Quantity
25,000
Distribution
Worldwide distribution: USA, UK, Netherlands, Canada, Australia, Denmark, Germany, Greece, Hong Kong, Israel, Italy, Philippines, Portugal, Singapore, South Africa, Spain Switzerland, Turkey and UAE.
Lot, serial or model codes
All units of the Catalog Numbers.
510(k) numbers
not on file
PMA numbers
["P040001"]
Product code
NQO
Firm FEI
3004904811
Firm city
Sunnyvale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register