Skip to the content

FDA Medical Device Recalls (openFDA)

Angiodynamics, Inc.: DURAMAX (TM) Hemodialysis Catheters, Stacked Tip 32 cm Straight Basic Kit, Catalog Number/REF 10302805, 510 (k) 080400, Sterile --- Manufactured in USA by AngioDynamics, Inc. 603 Queensbury Avenue, Queensbury, NY 12804

Recall number
Z-1764-2010
Recalling firm
Angiodynamics, Inc.
Product
DURAMAX (TM) Hemodialysis Catheters, Stacked Tip 32 cm Straight Basic Kit, Catalog Number/REF 10302805, 510 (k) 080400, Sterile --- Manufactured in USA by AngioDynamics, Inc. 603 Queensbury Avenue, Queensbury, NY 12804
Status
Terminated
Initiated
2010-01-08
Posted
2010-06-04
Terminated
2012-07-25
Reason
The product has the potential for a cross-lumen leak. One of the two leak testers that are used to perform the leak test on the DuraMax Catheters was found to be faulty. The tester was not able to detect cross-lumen leaks. Both testers were used during the manufacturing 100% leak test.
Root cause
Other
Action
Customers were sent URGENT MEDICAL DEVICE RECALL LETTERS (dated 1/07/10) via Certified Mail, Return Receipt Requested. Customers were instructed to return all recalled product. A return authorization number was assigned to each customer in the recall notification letter and a UPS account was provided for return freight. A reply form was provided in the recall notification letter. Questions were directed to the company sales representatives or to Customer Service at 800-772-6446.
Quantity
240
Distribution
Nationwide and St. Thomas, VI
Lot, serial or model codes
Lots 978913, 980648, 981115, C68440, C68441
510(k) numbers
not on file
PMA numbers
not on file
Product code
MSD
Firm FEI
1319211
Firm city
Queensbury
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register