Skip to the content

FDA Medical Device Recalls (openFDA)

Focus Diagnostics Inc: HerpeSelect 1 and 2 Immunoblot IgG Kit, IB0900G Intended for qualitatively detecting the presence or absence of human IgG class antibodies to HSV-1 and HSV-2 in human sera.

Recall number
Z-1764-2008
Recalling firm
Focus Diagnostics Inc
Product
HerpeSelect 1 and 2 Immunoblot IgG Kit, IB0900G Intended for qualitatively detecting the presence or absence of human IgG class antibodies to HSV-1 and HSV-2 in human sera.
Status
Terminated
Initiated
2007-07-06
Posted
2008-08-17
Terminated
2009-06-25
Reason
A small percentage of samples tested in specific lots, the HSV common antigen band appears lighter (less reactive) than the reading control band. Reduced reactivity of the common antigen band may lead to inaccurate interpretation of a sample as negative.
Root cause
Process design
Action
Focus notified all clients which received the recalled lots starting on July 6, 2007 via an Important Medical Device Information letter by fax or email. The letter explains how to identify suspect results and the steps to be taken if suspect results are obtained. If customers find the interpretation of results with the indicated kit lots to be problematic, customers have the option of having their kits replaced at no charge within 90 days of receiving the notification.
Quantity
Lot Code 063169 - 97 Kits, Lot Code 063354 - 16 Kits, Lot Code 063355 - 74 Kits
Distribution
Worldwide Distribution --- USA including states of OH, CA, CT, IL, NY, and countries of Barbados, Canada, Germany, Israel, Slovakia, France, and England.
Lot, serial or model codes
Lot Codes 063169 (Expires December 2007), 063354 (Expires October 2007), 063355 (Expires July 2007)
510(k) numbers
K000238
PMA numbers
not on file
Product code
LGC
Firm FEI
2023365
Firm city
Cypress
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this source