Focus Diagnostics Inc: HerpeSelect 1 and 2 Immunoblot IgG Kit, IB0900G Intended for qualitatively detecting the presence or absence of human IgG class antibodies to HSV-1 and HSV-2 in human sera.
- Recall number
- Z-1764-2008
- Recalling firm
- Focus Diagnostics Inc
- Product
- HerpeSelect 1 and 2 Immunoblot IgG Kit, IB0900G Intended for qualitatively detecting the presence or absence of human IgG class antibodies to HSV-1 and HSV-2 in human sera.
- Status
- Terminated
- Initiated
- 2007-07-06
- Posted
- 2008-08-17
- Terminated
- 2009-06-25
- Reason
- A small percentage of samples tested in specific lots, the HSV common antigen band appears lighter (less reactive) than the reading control band. Reduced reactivity of the common antigen band may lead to inaccurate interpretation of a sample as negative.
- Root cause
- Process design
- Action
- Focus notified all clients which received the recalled lots starting on July 6, 2007 via an Important Medical Device Information letter by fax or email. The letter explains how to identify suspect results and the steps to be taken if suspect results are obtained. If customers find the interpretation of results with the indicated kit lots to be problematic, customers have the option of having their kits replaced at no charge within 90 days of receiving the notification.
- Quantity
- Lot Code 063169 - 97 Kits, Lot Code 063354 - 16 Kits, Lot Code 063355 - 74 Kits
- Distribution
- Worldwide Distribution --- USA including states of OH, CA, CT, IL, NY, and countries of Barbados, Canada, Germany, Israel, Slovakia, France, and England.
- Lot, serial or model codes
- Lot Codes 063169 (Expires December 2007), 063354 (Expires October 2007), 063355 (Expires July 2007)
- 510(k) numbers
- K000238
- PMA numbers
- not on file
- Product code
- LGC
- Firm FEI
- 2023365
- Firm city
- Cypress
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28