Roche Diagnostics Corporation: Cobas 8100 bi-directional reformatter (BRF) module with Software Version 02-xx
- Recall number
- Z-1763-2017
- Recalling firm
- Roche Diagnostics Corporation
- Product
- Cobas 8100 bi-directional reformatter (BRF) module with Software Version 02-xx
- Status
- Terminated
- Initiated
- 2017-03-01
- Posted
- 2017-03-23
- Terminated
- 2018-07-03
- Reason
- "Roche Diagnostics Corp. initiated a voluntary correction because a rack crash may occur on the Cobas 8100 bi-directional reformatter (BRF) and uni-directional reformatter (URF) modules with software version 02-xx, only when rack buffering is activated and the rack buffer is completely full. This issue can lead to sample spillage, posing a potential risk to operators / laboratory staff due to exposure to potentially infectious material, cross-contamination of samples in the affected racks and erroneous results due to sample carry-over."
- Root cause
- Under Investigation by firm
- Action
- Roche initiated a recall on 03/01/2017, notices were issued via UPS to customers and instructed customers to do the following: Roche is developing software version 03-01 to resolve this issue, and a Roche Representative will be contacting you in the near future to schedule an update. Until this software update is available, the rack buffering option for the BRF and URF modules must be disabled. If you are an authorized cobas 8100 automated workflow series operator, follow the steps outlined below to disable rack buffering. If your facility does not have an authorized cobas 8100 automated workflow series operator, please contact the Roche Support Network Customer Support Center at 1-800-428-2336 for further steps. If you have a major spill, set the affected module to offline and contact the Roche Support Network Customer Support Center at 1-800-428-2336. " If you are an authorized cobas 8100 automated workflow series operator, follow the steps outlined in this Urgent Medical Device Correction (UMDC) to disable rack buffering. " If your facility does not have an authorized cobas 8100 automated workflow series operator, please contact the Roche Support Network Customer Support Center at 1-800-428-2336 for further steps. " Complete the attached faxback form and fax it to 1-888-345-0024. " File this Urgent Medical Device Correction (UMDC) for future reference. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in this UMDC.
- Quantity
- 45 Units
- Distribution
- Domestic: MA, IN, AL, OH, MO, NE, LA, CA, IA, NJ, AR, PA, SC, IL, TX, and MI. Foreign: None
- Lot, serial or model codes
- Cobas 8100 bi-directional reformatter (BRF) module with Software Version 02-xx Catalog Number: 743995400
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- JQP
- Firm FEI
- 1823260
- Firm city
- Indianapolis
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28