Skip to the content

FDA Medical Device Recalls (openFDA)

Teleflex Medical: Hudson RCI Sheridan: 1) EZ-ENDO 6.0 mm, Product Code 5-22512 2) EZ-ENDO 6.5 mm, Product Code 5-22513 3) EZ-ENDO 7.0 mm, Product Code 5-22514 4) EZ-ENDO 7.5 mm, Product Code 5-22515 5) EZ-ENDO 8.0 mm, Product Code 5-22516 6) EZ-ENDO 8.5 mm, Product Code 5-22517 Product Usage: Tracheal tube/airway management

Recall number
Z-1761-2019
Recalling firm
Teleflex Medical
Product
Hudson RCI Sheridan: 1) EZ-ENDO 6.0 mm, Product Code 5-22512 2) EZ-ENDO 6.5 mm, Product Code 5-22513 3) EZ-ENDO 7.0 mm, Product Code 5-22514 4) EZ-ENDO 7.5 mm, Product Code 5-22515 5) EZ-ENDO 8.0 mm, Product Code 5-22516 6) EZ-ENDO 8.5 mm, Product Code 5-22517 Product Usage: Tracheal tube/airway management
Status
Terminated
Initiated
2019-05-24
Posted
2019-06-19
Terminated
2022-02-15
Reason
This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.
Root cause
Process control
Action
Teleflex sent an Urgent Medical Device Recall letter dated May 24, 2019 to customers. The letter identified the affected product, problem and actions to be taken. The letter directed customers to discontinue use, quarantine the product, and contact Teleflex for return of the product. Distributors were directed to notify their customers and request the customer return the recalled products to them for consolidation and subsequent return to Teleflex.. For questions contact your local sales representative or Customer Service at 1-866-396-2111
Quantity
55312 units
Distribution
Worldwide - US Nationwide Distribution
Lot, serial or model codes
1) EZ-ENDO 6.0 mm, Product Code 5-22512 Lot Numbers: 73K1600152 73M1600163 73A1700280 73C1700567 73E1700496 73F1700403 73H1700347 73M1700272 73E1800401 73E1800575 73F1800354  2) EZ-ENDO 6.5 mm, Product Code 5-22513 Lot Numbers: 73J1600766 73B1700155 73C1700568 73F1700404 73G1700432 73H1700348 73E1800576 73F1800355  3) EZ-ENDO 7.0 mm, Product Code 5-22514 Lot Numbers: 73A1700118 73C1700569 73E1700722 73F1700175 73F1700405 73G1700433 73H1700349 73H1700494 73H1700711 73J1700143 73K1700081 73K1700610 73L1700181 73M1700273 73D1800653 73F1800562 73F1800720 73H1800450 73J1800531 73K1800328 73K1800786  4) EZ-ENDO 7.5 mm, Product Code 5-22515 Lot Numbers: 73B1700316 73C1700262 73C1700418 73E1700723 73E1700104 73F1700176 73F1700406 73G1700434 73H1700350 73J1700302 73J1700497 73K1700747 73M1700274 73A1800068 73A1800683 73B1800502 73C1800201 73D1800154 73E1800402 73F1800565 73G1800591 73H1800451 73H1800686 73J1800113 73K1800736  5) EZ-ENDO 8.0 mm, Product Code 5-22516 Lot Numbers: 73K1600594 73A1700282 73B1700158 73B1700317 73C1700570 73F1700036 73F1700177 73F1700407 73G1700614 73H1700495 73J1700487 73K1700082 73A1800684 73B1800503 73C1800371 73E1800577 73G1800232 73J1800532 73K1800327  6) EZ-ENDO 8.5 mm, Product Code 5-22517 Lot Numbers: 73J1600767 73C1800202
510(k) numbers
["K822082"]
PMA numbers
not on file
Product code
BTR
Firm FEI
3005747797
Firm city
Morrisville
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register