FDA Medical Device Recalls (openFDA)
Bard Peripheral Vascular Inc: POWERPORT(R) ClearVUE(R) isp with Smooth Septum, 6F ChronoFlex(TM), Polyurethane Catheter, REF 1666362
- Recall number
- Z-1760-2018
- Recalling firm
- Bard Peripheral Vascular Inc
- Product
- POWERPORT(R) ClearVUE(R) isp with Smooth Septum, 6F ChronoFlex(TM), Polyurethane Catheter, REF 1666362
- Status
- Completed
- Initiated
- 2018-02-08
- Posted
- not on file
- Terminated
- not on file
- Reason
- The Patient ID Card packaged with the POWERPORT ClearVUE Implantable Port with Smooth Septum contains incorrect port flushing instructions.
- Root cause
- Error in labeling
- Action
- The firm initiated the recall by letter on 02/08/2018. The letter explained the problem and the products involved and requested the following actions: "Please check all inventory locations within your institution for affected product code / lot number combinations listed in the field correction notice. If you have further distributed any of the product code / lot numbers, please immediately contact that location, advise them of this field correction notice, and forward these instructions. For any identified product on your shelves, please replace the products Patient ID Card with the corrected Patient ID Card enclosed with this field correction notice. Enclosed with this field correction notice is a Patient Communication Packet that contains one patient letter, one corrected Patient ID Card, and one POWERPORT Patient Guide. Please contact all patients that have received any product listed in this field correction notice and provide them with the Patient Communication Packet."
- Quantity
- 359 units
- Distribution
- Nationwide
- Lot, serial or model codes
- Lot Number Unique Device Identifier (UDI) REAN0063 (01)00801741110726(17)170628(10)REAN0063 REAN1123 (01)00801741110726(17)170628(10)REAN1123 REAP0467 (01)00801741110726(17)170728(10)REAP0467 REAY1733 (01)00801741110726(17)180331(10)REAY1733 REBS1266 (01)00801741110726(17)180731(10)REBS1266 REBV1731 (01)00801741110726(17)181231(10)REBV1731 REBX1626 (01)00801741110726(17)190331(10)REBX1626 REBY0777 (01)00801741110726(17)190430(10)REBY0777
- 510(k) numbers
- ["K063377"]
- PMA numbers
- not on file
- Product code
- LJT
- Firm FEI
- 2020394
- Firm city
- Tempe
- Firm state
- AZ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28