Skip to the content

FDA Medical Device Recalls (openFDA)

ConMed Corporation: HYFRECATOR 2000 HANDPIECE SHEATH, REF/Part Number 7-796-19, Rx ONLY, STERILE, Single Patient Use -- CONMED CORPORATION, 525 FRENCH ROAD UTICA, NY 13502. --- There is no problem with the Hyfrecator 2000 Electrosurgical Units or with cases of P/N 7-796-19 sold in cases as P/N 7-796-19BX or P/N 7-796-19CS.

Recall number
Z-1760-2017
Recalling firm
ConMed Corporation
Product
HYFRECATOR 2000 HANDPIECE SHEATH, REF/Part Number 7-796-19, Rx ONLY, STERILE, Single Patient Use -- CONMED CORPORATION, 525 FRENCH ROAD UTICA, NY 13502. --- There is no problem with the Hyfrecator 2000 Electrosurgical Units or with cases of P/N 7-796-19 sold in cases as P/N 7-796-19BX or P/N 7-796-19CS.
Status
Terminated
Initiated
2017-02-28
Posted
not on file
Terminated
not on file
Reason
For over a year, the Accessory Packages in the Hyfrecator 2000 ESU cartons were packaged in error by the manufacturer with non-sterile P/N 7-796-19 Handpiece Sheaths in pouches that were labeled as Sterile. This could cause an end user to use a non-sterile sheath as if it were Sterile.
Root cause
Mixed-up of materials/components
Action
ConMed sent an URGENT: MEDICAL DEVICE RECALL Notices (dated 2/28/2017)to all affected US customers by USPS First Class Mail. International letters are in the process of translation and are scheduled to be sent by March 15, 2017. Customers were instructed to review their inventory for any of the affected product and were asked to complete the business reply form and return it to ConMed Corporation, 525 French Road, Utica, New York 13502. Customers with questions or requests should contact the Field Action Support Team at 1-800-448-6506, fax to 315-624-3225, or email hyfr2000@conmed.com. For questions regarding this recall call 315-624-3533.
Quantity
not on file
Distribution
Worldwide Distribution - US (nationwide) Internationally to Australia, Canada, Germany, Italy, UK, Great Britain, and Ireland.
Lot, serial or model codes
Affected Lot Numbers: 201511234, 201512074, 201512094, 201601064, 201603284
510(k) numbers
["K963088"]
PMA numbers
not on file
Product code
GEI
Firm FEI
3001451571
Firm city
Utica
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register