Teleflex Medical: Hudson RCI Sheridan LTS, Product Code 5-11112 Product Usage: Tracheal tube/airway management
- Recall number
- Z-1759-2019
- Recalling firm
- Teleflex Medical
- Product
- Hudson RCI Sheridan LTS, Product Code 5-11112 Product Usage: Tracheal tube/airway management
- Status
- Terminated
- Initiated
- 2019-05-24
- Posted
- 2019-06-19
- Terminated
- 2022-02-15
- Reason
- This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.
- Root cause
- Process control
- Action
- Teleflex sent an Urgent Medical Device Recall letter dated May 24, 2019 to customers. The letter identified the affected product, problem and actions to be taken. The letter directed customers to discontinue use, quarantine the product, and contact Teleflex for return of the product. Distributors were directed to notify their customers and request the customer return the recalled products to them for consolidation and subsequent return to Teleflex.. For questions contact your local sales representative or Customer Service at 1-866-396-2111
- Quantity
- 3483 units
- Distribution
- Worldwide - US Nationwide Distribution
- Lot, serial or model codes
- Lot Numbers: 73K1600369, 73L1600011, 73A1700646, 73C1700238, 73D1700625, 73E1700475, 73F1700387, 73H1700041, 73H1700483, 73J1700292, 73L1700503, 73M1700453, 73A1800671, 73B1800350, 73D1800138, 73F1800174, 73G1800126, 73G1800806, 73J1800308, 73K1800173, 73L1800225, 73K1800172
- 510(k) numbers
- ["K822082"]
- PMA numbers
- not on file
- Product code
- BTR
- Firm FEI
- 3005747797
- Firm city
- Morrisville
- Firm state
- NC
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z