FDA Medical Device Recalls (openFDA)
Philips Ultrasound, Inc.: 5000 Compact Series Ultrasound Systems, Product/REF: 5500 CV/795141, 5500 P/795143, 5500 W/795144, 5500 G/795145, and Product/REF: 5300 W/795146, 5300 P/795147, 5300 G/795148, used with Deluxe and Premium Carts that come with multiport adapters.
- Recall number
- Z-1757-2024
- Recalling firm
- Philips Ultrasound, Inc.
- Product
- 5000 Compact Series Ultrasound Systems, Product/REF: 5500 CV/795141, 5500 P/795143, 5500 W/795144, 5500 G/795145, and Product/REF: 5300 W/795146, 5300 P/795147, 5300 G/795148, used with Deluxe and Premium Carts that come with multiport adapters.
- Status
- Open, Classified
- Initiated
- 2024-03-23
- Posted
- 2024-05-03
- Terminated
- not on file
- Reason
- Ultrasound system with: 1) transesophageal echocardiography transducer (TEE) connected/not selected 2) 1, 2 non-TEE connected 3) Imaging with non-TEE 4) Selected/not selected imaging transducer removed 5) Transducer connected 6) Non-TEE imaging continues, Then though TEE is unselected, it may have power, no temperature monitoring/control, and may result in temperature above limit and patient burns
- Root cause
- Component change control
- Action
- On 3/23/24, correction notices were distributed to customers who were advised of the following two ways to avoid the condition: Option 1: To prevent the issue from occurring, when using the TEE transducers, connect directly to the Compact 5000 system using the port on the side of the system as shown in Figure 3 below. The issue is associated with the MPA, therefore by using the port on the Compact 5000 directly, the issue is avoided. Option 2: Whenever adding a transducer to, or removing a transducer from the MPA, disconnect and reconnect the MPA connector on the Compact 5000 system connector port, followed by a transducer selection on the touch screen. By performing this step when adding or removing a transducer, the monitoring and control systems are re-initialized. Circulate this URGENT Medical Device Correction Letter to all users of this device within your organization and forward to any organization where potentially affected devices have been transferred so that they are aware of the issue. In addition, customers were asked to complete and return the response form ultrasound.corrections@philips.com A firm representative will contact you to schedule a time for a Field Service Engineer to visit and implement the solution to resolve the issue For further information or support concerning this issue, please contact your local firm representative: For North America, contact the Customer Care Solutions Center (1-800-722-9377).
- Quantity
- 363
- Distribution
- Worldwide - US Nationwide distribution including in the states of CA, CO, CT, FL, GA, IL, IN, KS, KY, MA, MD, ME, MI, NC, NH, NM, NY, OH, OK, PA, PR, SC, TN, TX, UT, VA and the countries of Australia, Austria, Bahrain, Brazil, Bulgaria, Canada, Croatia, Czech Republic, Denmark, Finland, France, Germany, Italy, Japan, Lithuania, Netherlands, New Zealand, Oman, Poland, Romania, Slovenia, South Africa, Spain, Sweden, Switzerland, United Kingdom, United Arab Emirates.
- Lot, serial or model codes
- All Compact 5000 series systems manufactured on or before 20-Dec-2023 and purchased with Deluxe and Premium Cart options. REF/UDI-DI/Serial Numbers: 505500 CV/795141/00884838116566/USN23P0684, USN23P0623, USN23P0624, USN23P0685, USN23P0720, USN23P0615, USN23P0627, USN23P0611, USN23P0683, USN23P0682, USN23P0625, USN23P0173, USN23P0164, US823P0231, USN23P0613, USN23P0172, USN23P0721, USN23P0410, USN23P0612, USN23P0614, USN23P0171, US623P0141, US823P0233, US423P0844, US523P1060, US523P1061, US523P1750, US523P2536, US523P2537, US523P2672, US523P2673, US423P0842, US623P0140, US423P0759, US623P0142, US623P0230, US623P0485, US623P0486, US623P0740, US623P0790, US623P0791, US623P0792, US623P0866, US623P0867, US423P0843, US623P0966, US423P0672, US423P0673, US423P0674, US423P0712, US423P0757, US823P0132, US823P0232, US823P0318, US823P0319, US823P0320, US823P0321, US823P0322, US823P0507, US823P0508, US823P1857, US923P1161, US823P0129, US723P1919, US623P1072, US623P1403, US623P1483, US623P1606, US623P1607, US623P1608, US623P1609, US823P0131, US823P0029, US823P0128, US823P0127, US823P0033, US823P0032, US823P0030, US823P0028, US723P2256, US723P2154, US723P2153, US723P2152, US823P0130, US823P0031, USO23P2173, USO23P2065, US723P1242, US723P1243, US723P1342, US723P1343, US723P1241, US723P1240, USN23P0333, USN23P0332, US623P1450, US723P1766, USO23P0157, USO23P0156, USO23P0059, USO23P0058, US823P1743, US823P1744, US823P1745, US723P1582, US723P1872, US723P1583, USO23P2274, USO23P2064, USO23P2171, US623P1739, US723P1239, US723P1111, USN23P0330, USN23P0331, US923P0205, USN23P0051, US823P1767, US923P0207, US823P1770, US823P1769, US823P1251, US823P1229, US723P0393, US923P0421, USO23P2069; 5500 P/795143/00884838116573/USN23P0048, USN23P0050, USO23P2276, USO23P2275, US523P2372, US523P2538, US523P2406, US523P2405, US523P2374, US523P2373, US523P1837, US523P1838, US623P0741, US523P1839, US823P0744, US823P0745, US723P2257, US823P1030, US823P0877, US823P0548, US823P0547, US823P0546, US823P0511, US823P0509, US823P0317, US723P2258, US823P0510, US623P1486, US623P1485, US623P1741, US623P0969, US623P0968, US723P0809, US823P2055, US823P2054, US723P1584, US723P1472, US723P1471, USN23P0334, US823P2053, US723P1917, US923P0381, US823P1254, US723P0182, US723P0063, US623P2118, USO23P0171, US723P1918, US823P1250, USO23P2174, US723P0171, US723P0172, US623P2117, US723P0061, US723P0062, US723P0170, USN23P0753, US723P1809. US922P0888. US723P0596, US623P1738, US723P1112, US723P1113, US723P1767; 5500 W/795144/00884838116580/USN23P0628, USN23P0411, USN23P0174, USN23P0166, US523P2404, US623P0603, US623P0601, US423P0994, US423P0993, US423P0995, US623P0602, US623P2004, US623P1075, US723P0174, US723P0173, US623P2109, US623P2005, US922P0887, US623P1740; 5500 G/795145/00884838116597/USN23P0165, USN23P0053, USO23P1901, USO23P2068, USN23P0686, USN23P0414, USN23P0413, USN23P0412, USN23P0335, US523P2674, US523P2800, US523P2798, US523P2799, US623P0318, US623P0395, US623P0394, US623P0393, US623P0319, US623P0229, US623P0228, US623P0019, US623P0018, US623P0017, US623P0320, US523P2152, US523P2151, US523P2014, US523P2013, US523P2012, US523P1749, US523P1440, US523P1439, US823P0984, US823P0742, US823P0743, US823P0730, US823P0879, US823P1084, US823P0550, US823P0512, US823P0732, US723P2261, US723P2260, US723P2259, US623P1281, US623P1610, US623P1404, US623P1280, US623P1713, US823P0549, US823P1853, US823P1855, US823P1856, US823P1854, US723P1473, US723P1474, US723P1344, US723P1346, US723P1345, US723P1475, US823P1676, US823P0505, US723P2155, US723P2156, US823P1652, USN23P0626, US823P1678, US723P2157, USN23P0170, US723P1765, US623P2119, US823P1230, US723P0183, USO23P0172, USO23P0173, USO23P0174, US823P1677, US723P1763, US723P0064, US823P1747, US823P1746, US823P0741, USO23P2175, US723P1585, US923P0204, US723P0390, US723P0185, US723P0184, US723P0391, US723P0259, US723P0186, US723P0392, US723P0260, US723P0258, US723P0257, US723P0256, US923P0419, USN23P0049, US423P0714, US523P0946, US223P0651, US723P0599, US723P0598, US723P0597, US823P0880, US223P0650, US623P1863, US623P1864, USO23P2279, US723P1114, US723P1115, US723P1914, US723P1915, US723P1916, US623P2000, USO23P1805, US823P1768, USO23P2277, US823P2052, US723P0743, US723P0744, US823P2051, US723P0993, US723P0992, US923P0206, US823P1232, US923P0420, US723P1764, US823P1231, US923P0208; 5300 W/795146/00884838116603/US922P0893, USN23P0616, USN23P0177; 5300 P/795147/00884838116610/USN23P0724, US922P0892, USN23P0415, US723P0850, US823P0836, US723P0810, US623P0908, US623P2124, US623P2123; 5300 G/795148/00884838116627/US922P0891, US823P0230, US823P0228, US623P0484, US823P0229, USO23P2070, USO23P2176, US723P1586, USN23P0175, USN23P0176, US723P0600, USN23P0052
- 510(k) numbers
- ["K222648"]
- PMA numbers
- not on file
- Product code
- IYN
- Firm FEI
- 3019216
- Firm city
- Bothell
- Firm state
- WA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28