Philips Ultrasound, Inc.: 5000 Compact Series Ultrasound Systems, Product/REF: 5500 CV/795141, 5500 P/795143, 5500 W/795144, 5500 G/795145, and Product/REF: 5300 W/795146, 5300 P/795147, 5300 G/795148, used with Deluxe and Premium Carts that come with multiport adapters.

Recall number
Z-1757-2024
Recalling firm
Philips Ultrasound, Inc.
Product
5000 Compact Series Ultrasound Systems, Product/REF: 5500 CV/795141, 5500 P/795143, 5500 W/795144, 5500 G/795145, and Product/REF: 5300 W/795146, 5300 P/795147, 5300 G/795148, used with Deluxe and Premium Carts that come with multiport adapters.
Status
Open, Classified
Initiated
2024-03-23
Posted
2024-05-03
Terminated
not on file
Reason
Ultrasound system with: 1) transesophageal echocardiography transducer (TEE) connected/not selected 2) 1, 2 non-TEE connected 3) Imaging with non-TEE 4) Selected/not selected imaging transducer removed 5) Transducer connected 6) Non-TEE imaging continues, Then though TEE is unselected, it may have power, no temperature monitoring/control, and may result in temperature above limit and patient burns
Root cause
Component change control
Action
On 3/23/24, correction notices were distributed to customers who were advised of the following two ways to avoid the condition: Option 1: To prevent the issue from occurring, when using the TEE transducers, connect directly to the Compact 5000 system using the port on the side of the system as shown in Figure 3 below. The issue is associated with the MPA, therefore by using the port on the Compact 5000 directly, the issue is avoided. Option 2: Whenever adding a transducer to, or removing a transducer from the MPA, disconnect and reconnect the MPA connector on the Compact 5000 system connector port, followed by a transducer selection on the touch screen. By performing this step when adding or removing a transducer, the monitoring and control systems are re-initialized. Circulate this URGENT Medical Device Correction Letter to all users of this device within your organization and forward to any organization where potentially affected devices have been transferred so that they are aware of the issue. In addition, customers were asked to complete and return the response form ultrasound.corrections@philips.com A firm representative will contact you to schedule a time for a Field Service Engineer to visit and implement the solution to resolve the issue For further information or support concerning this issue, please contact your local firm representative: For North America, contact the Customer Care Solutions Center (1-800-722-9377).
Quantity
363
Distribution
Worldwide - US Nationwide distribution including in the states of CA, CO, CT, FL, GA, IL, IN, KS, KY, MA, MD, ME, MI, NC, NH, NM, NY, OH, OK, PA, PR, SC, TN, TX, UT, VA and the countries of Australia, Austria, Bahrain, Brazil, Bulgaria, Canada, Croatia, Czech Republic, Denmark, Finland, France, Germany, Italy, Japan, Lithuania, Netherlands, New Zealand, Oman, Poland, Romania, Slovenia, South Africa, Spain, Sweden, Switzerland, United Kingdom, United Arab Emirates.
Lot, serial or model codes
All Compact 5000 series systems manufactured on or before 20-Dec-2023 and purchased with Deluxe and Premium Cart options. REF/UDI-DI/Serial Numbers: 505500 CV/795141/00884838116566/USN23P0684, USN23P0623, USN23P0624, USN23P0685, USN23P0720, USN23P0615, USN23P0627, USN23P0611, USN23P0683, USN23P0682, USN23P0625, USN23P0173, USN23P0164, US823P0231, USN23P0613, USN23P0172, USN23P0721, USN23P0410, USN23P0612, USN23P0614, USN23P0171, US623P0141, US823P0233, US423P0844, US523P1060, US523P1061, US523P1750, US523P2536, US523P2537, US523P2672, US523P2673, US423P0842, US623P0140, US423P0759, US623P0142, US623P0230, US623P0485, US623P0486, US623P0740, US623P0790, US623P0791, US623P0792, US623P0866, US623P0867, US423P0843, US623P0966, US423P0672, US423P0673, US423P0674, US423P0712, US423P0757, US823P0132, US823P0232, US823P0318, US823P0319, US823P0320, US823P0321, US823P0322, US823P0507, US823P0508, US823P1857, US923P1161, US823P0129, US723P1919, US623P1072, US623P1403, US623P1483, US623P1606, US623P1607, US623P1608, US623P1609, US823P0131, US823P0029, US823P0128, US823P0127, US823P0033, US823P0032, US823P0030, US823P0028, US723P2256, US723P2154, US723P2153, US723P2152, US823P0130, US823P0031, USO23P2173, USO23P2065, US723P1242, US723P1243, US723P1342, US723P1343, US723P1241, US723P1240, USN23P0333, USN23P0332, US623P1450, US723P1766, USO23P0157, USO23P0156, USO23P0059, USO23P0058, US823P1743, US823P1744, US823P1745, US723P1582, US723P1872, US723P1583, USO23P2274, USO23P2064, USO23P2171, US623P1739, US723P1239, US723P1111, USN23P0330, USN23P0331, US923P0205, USN23P0051, US823P1767, US923P0207, US823P1770, US823P1769, US823P1251, US823P1229, US723P0393, US923P0421, USO23P2069; 5500 P/795143/00884838116573/USN23P0048, USN23P0050, USO23P2276, USO23P2275, US523P2372, US523P2538, US523P2406, US523P2405, US523P2374, US523P2373, US523P1837, US523P1838, US623P0741, US523P1839, US823P0744, US823P0745, US723P2257, US823P1030, US823P0877, US823P0548, US823P0547, US823P0546, US823P0511, US823P0509, US823P0317, US723P2258, US823P0510, US623P1486, US623P1485, US623P1741, US623P0969, US623P0968, US723P0809, US823P2055, US823P2054, US723P1584, US723P1472, US723P1471, USN23P0334, US823P2053, US723P1917, US923P0381, US823P1254, US723P0182, US723P0063, US623P2118, USO23P0171, US723P1918, US823P1250, USO23P2174, US723P0171, US723P0172, US623P2117, US723P0061, US723P0062, US723P0170, USN23P0753, US723P1809. US922P0888. US723P0596, US623P1738, US723P1112, US723P1113, US723P1767; 5500 W/795144/00884838116580/USN23P0628, USN23P0411, USN23P0174, USN23P0166, US523P2404, US623P0603, US623P0601, US423P0994, US423P0993, US423P0995, US623P0602, US623P2004, US623P1075, US723P0174, US723P0173, US623P2109, US623P2005, US922P0887, US623P1740; 5500 G/795145/00884838116597/USN23P0165, USN23P0053, USO23P1901, USO23P2068, USN23P0686, USN23P0414, USN23P0413, USN23P0412, USN23P0335, US523P2674, US523P2800, US523P2798, US523P2799, US623P0318, US623P0395, US623P0394, US623P0393, US623P0319, US623P0229, US623P0228, US623P0019, US623P0018, US623P0017, US623P0320, US523P2152, US523P2151, US523P2014, US523P2013, US523P2012, US523P1749, US523P1440, US523P1439, US823P0984, US823P0742, US823P0743, US823P0730, US823P0879, US823P1084, US823P0550, US823P0512, US823P0732, US723P2261, US723P2260, US723P2259, US623P1281, US623P1610, US623P1404, US623P1280, US623P1713, US823P0549, US823P1853, US823P1855, US823P1856, US823P1854, US723P1473, US723P1474, US723P1344, US723P1346, US723P1345, US723P1475, US823P1676, US823P0505, US723P2155, US723P2156, US823P1652, USN23P0626, US823P1678, US723P2157, USN23P0170, US723P1765, US623P2119, US823P1230, US723P0183, USO23P0172, USO23P0173, USO23P0174, US823P1677, US723P1763, US723P0064, US823P1747, US823P1746, US823P0741, USO23P2175, US723P1585, US923P0204, US723P0390, US723P0185, US723P0184, US723P0391, US723P0259, US723P0186, US723P0392, US723P0260, US723P0258, US723P0257, US723P0256, US923P0419, USN23P0049, US423P0714, US523P0946, US223P0651, US723P0599, US723P0598, US723P0597, US823P0880, US223P0650, US623P1863, US623P1864, USO23P2279, US723P1114, US723P1115, US723P1914, US723P1915, US723P1916, US623P2000, USO23P1805, US823P1768, USO23P2277, US823P2052, US723P0743, US723P0744, US823P2051, US723P0993, US723P0992, US923P0206, US823P1232, US923P0420, US723P1764, US823P1231, US923P0208; 5300 W/795146/00884838116603/US922P0893, USN23P0616, USN23P0177; 5300 P/795147/00884838116610/USN23P0724, US922P0892, USN23P0415, US723P0850, US823P0836, US723P0810, US623P0908, US623P2124, US623P2123; 5300 G/795148/00884838116627/US922P0891, US823P0230, US823P0228, US623P0484, US823P0229, USO23P2070, USO23P2176, US723P1586, USN23P0175, USN23P0176, US723P0600, USN23P0052
510(k) numbers
["K222648"]
PMA numbers
not on file
Product code
IYN
Firm FEI
3019216
Firm city
Bothell
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register