FDA Medical Device Recalls (openFDA)
Baylis Medical Corp *: The Baylis LumbarCool Pain Management Kit consisting of: 1 LumbarCool Pain Management Probe, 3 LumbarCool Pain Management introducers, and 1 Pain Management Tube Kit. Model/Catalogue Number: LUK-17-150-4 The LumbarCool Pain Management System, a cooled radiofrequency pain management system, in combination with the Baylis Pain Management Generator-TD (PMG-TD), is indicated for use to create RF lesions in nervous tissue.
- Recall number
- Z-1755-2010
- Recalling firm
- Baylis Medical Corp *
- Product
- The Baylis LumbarCool Pain Management Kit consisting of: 1 LumbarCool Pain Management Probe, 3 LumbarCool Pain Management introducers, and 1 Pain Management Tube Kit. Model/Catalogue Number: LUK-17-150-4 The LumbarCool Pain Management System, a cooled radiofrequency pain management system, in combination with the Baylis Pain Management Generator-TD (PMG-TD), is indicated for use to create RF lesions in nervous tissue.
- Status
- Terminated
- Initiated
- 2010-05-12
- Posted
- 2010-06-03
- Terminated
- 2010-10-04
- Reason
- Name of the device reflected on the product packing sleeve is incorrect.
- Root cause
- Error in labeling
- Action
- The firm, Baylis Medical, notified US Distributor by email and sent a "URGENT Field Corrective Action Notice" dated May 12, 2010, to customer. The notifications described the product, problem and action to be taken by customer. The customers were given the follow options: Option 1: In order to prevent confusion in identification of the product before use, please immediately remove and dispose the package sleeve that covers the tray packaging of the LumbarCool Pain Management Kit. Option 2: Alternatively, you may exchange your kit from Lot LKFA160310 with a correctly labeled replacement LumbarCool Pain Management Kit by contacting Kimberly-Clark Health Care sales representative. The customers were ask to fill out the Field Corrective Action Acknowledgment and return it. Any question please call Baylis Sales Representative at 905-602-4875 and/or contact your local sales representative for any further information you may require.
- Quantity
- 58
- Distribution
- Nationwide distribution: TX
- Lot, serial or model codes
- Lot Number: LKFA160310
- 510(k) numbers
- ["K053082","K062937"]
- PMA numbers
- not on file
- Product code
- GXI
- Firm FEI
- 3005241199
- Firm city
- Montreal
- Firm state
- not on file
- Firm country
- Canada
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28