FDA Medical Device Recalls (openFDA)
LeMaitre Vascular, Inc.: Pruitt Occlusion Catheter. Catalog Numbers: 2103-36, 2103-46, 2103-56, E2103-36, E2103-46, E2103-56
- Recall number
- Z-1751-2025
- Recalling firm
- LeMaitre Vascular, Inc.
- Product
- Pruitt Occlusion Catheter. Catalog Numbers: 2103-36, 2103-46, 2103-56, E2103-36, E2103-46, E2103-56
- Status
- Open, Classified
- Initiated
- 2025-04-11
- Posted
- 2025-05-12
- Terminated
- not on file
- Reason
- During internal product testing, it was observed that some packages to have incomplete seals (sterile barrier).
- Root cause
- Packaging process control
- Action
- LeMaitre notified OUS consignees beginning on 04/11/2025, and US consignees on 04/23/2025 via email and letter. Consignees were instructed to not use the devices, immediately quarantine any on hand, and complete and return the response form. Once received a LeMaitre Vascular representative will contact customers to provide return instructions. If the products were transferred or further distributed, customers were instructed to notify those customers and provide a copy of the recall letter.
- Quantity
- 65,977 units
- Distribution
- Nationwide distribution. International distribution to ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, BULGARIA, CAMBODIA, CANADA, CHILE, COLOMBIA, COSTA RICA, CYPRUS, CZECHIA, DENMARK, DOMINICAN REPUBLIC, EGYPT, EL SALVADOR, ESTONIA, FINLAND, FRANCE, GEORGIA, GERMANY, GREECE, GUATEMALA, GUYANA, HONG KONG, HUNGARY, INDIA, INDONESIA, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KUWAIT, LATVIA, LEBANON, LITHUANIA, MALAYSIA, MALTA, MEXICO, MONGOLIA, MOROCCO, NETHERLANDS, NEW ZEALAND, NICARAGUA, NORTH MACEDONIA, NORWAY, OMAN, PANAMA, PARAGUAY, PHILIPPINES, POLAND, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TRINIDAD AND TOBAGO, TUNISIA, TURKIYE, UNITED ARAB EMIRATES, UNITED KINGDOM, VENEZUELA, VIETNAM.
- Lot, serial or model codes
- UDI-DI: 00840663101535, 00840663101542. Catalog Numbers: 2103-36, 2103-46, 2103-56, E2103-36, E2103-46, E2103-56. Lots not expired with the following prefix: POC, QPO, XPO.
- 510(k) numbers
- ["K811276"]
- PMA numbers
- not on file
- Product code
- DYG
- Firm FEI
- 1220948
- Firm city
- Burlington
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28