Skip to the content

FDA Medical Device Recalls (openFDA)

Biomet, Inc.: Comprehensive Humeral Tray Taper Extraction Pliers - Replacement Tips, Model Number 110028522 - Product Usage: Intended to be used in surgery to relieve pain and to replace the glenoid and humeral side of the shoulder joint in order to restore mobility in a grossly rotator cuff deficient joint with severe arthropathy or a previous joint replacement with a grossly rotator cuff deficient joint.

Recall number
Z-1750-2020
Recalling firm
Biomet, Inc.
Product
Comprehensive Humeral Tray Taper Extraction Pliers - Replacement Tips, Model Number 110028522 - Product Usage: Intended to be used in surgery to relieve pain and to replace the glenoid and humeral side of the shoulder joint in order to restore mobility in a grossly rotator cuff deficient joint with severe arthropathy or a previous joint replacement with a grossly rotator cuff deficient joint.
Status
Terminated
Initiated
2020-03-13
Posted
not on file
Terminated
2021-07-21
Reason
There were unknown bioburden levels prior to sterilization, which may affect sterility of the product.
Root cause
Under Investigation by firm
Action
URGENT MEDICAL DEVICE RECALL notification letters dated 3/13/20 were sent to customers. Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the field safety notice for hospitals and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form will be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com. Surgeon Responsibilities: 1. Review this notification for awareness of the contents. 2. There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule. 3. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 5. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Mon
Quantity
125
Distribution
Worldwide distribution - US Nationwide distribution including the states of AL, CO, KS, MA, MD, MI, MO, NH, NJ, and PA. The countries of Australia, Canada, India, and Netherlands.
Lot, serial or model codes
Lot Numbers: 494350 494350R 233080 364910 489000 079630 516230 742760 362370 465800 197900 175110 175100
510(k) numbers
["K080642"]
PMA numbers
not on file
Product code
PHX
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register