FUJIFILM Healthcare Americas Corporation: Surpria: Software Version V3.11, V3.22

Recall number
Z-1749-2022
Recalling firm
FUJIFILM Healthcare Americas Corporation
Product
Surpria: Software Version V3.11, V3.22
Status
Open, Classified
Initiated
2022-07-12
Posted
not on file
Terminated
not on file
Reason
Error 00003050 or 00003052 may occur during scanning, the captured images will not be reconstructed and the Raw Data will not be displayed in the work list.result, may result in the need for additional scanning to capture the images properly for reconstruction, thereby exposing the patient to additional radiation
Root cause
Under Investigation by firm
Action
FujiFilm issued Urgent Medical Device Letter were sent via email on July 12, 2022, consignees without an email contact were notified via FedEx on July 12, 2022 and July 15, 2022. Letter states reason for recall, health risk and action to take: Until your upgrade is scheduled and completed, you can continue to safely use your CT system by following these additional instructions: 1. When entering alphabetic characters in the patient information, please enter them in capital letters (e.g. ready READY ) 2. Set Auto Capital Letter to ON in the Preference Settings window. This will convert all text to capital letters for certain patient information fields. 3. Please register the following information in capital letters, as Auto Capital Letter does not apply to the following fields: a. (0008,0050) Accession Number b. (0008,0090) Referring Physician s Name c. (0010,0010) Patient s Name d. (0010,0020) Patient ID e. (0010,4000) Patient Comments f. (0032,1032) Requesting Physician g. (0032,1033) Requesting Service h. (0032,1060) Requested Procedure Description i. (0032,1070) Requested Contrast Agent j. (0040,0006) Scheduled Performing Physician s Name Please complete and return the Field Action Verification Form attached on page 5 of this communication. Responding with the information requested is essential for ensuring appropriate action is taken. Fujifilm regrets any inconvenience this action may cause and appreciates your understanding and cooperation. Please contact our field action coordinator, Jeffrey Wan, via phone at 201-675-8947 or email at jeffrey.wan@fujifilm.com
Quantity
20 units
Distribution
US Nationwide Distribution
Lot, serial or model codes
UDI: (01)04580292766601(11)200130(21)W5067(250)G560045319 (01)04580292766601(11)200220(21)W5068(250)G560048119 (01)04580292766601(11)200902(21)W5069(250)G560071519 (01)04580292766601(11)201210(21)W5070(250)G560075519 (01)04580292766601(11)201225(21)W5071(250)G1Z0040620 (01)04580292766601(11)210218(21)W5072(250)G1Z0063520 (01)04580292766601(11)210512(21)W5073(250)G1Z0244920 (01)04580292766601(11)210519(21)W5074(250)G1Z0048121 (01)04573596200520(11)210818(21)W5075(250)G1Z0083221 (01)04573596200520(11)210903(21)W5076(250)G1Z0094621 (01)04573596200520(11)210903(21)W5077(250)G1Z0094521 (01)04573596200520(11)210903(21)W5077(250)G1Z0094521 (01)04573596200520(11)211018(21)W5079(250)G1Z0123621 (01)04573596200520(11)211115(21)W5080(250)G1Z0190221 (01)04573596200520(11)211122(21)W5081(250)G1Z0190021 (01)04573596200520(11)211122(21)W5082(250)G1Z0190121 (01)04573596200520(11)211227(21)W5083(250)G1Z0190321 (01)04573596200520(11)211228(21)W5084(250)G1Z0210821 (01)04573596200520(11)220209(21)W5085(250)G1Z0210921 (01)04573596200520(11)220215(21)W5086(250)G1Z0211021 Serial Numbers: G560045319 G560048119 G560071519 G560075519 G1Z0040620 G1Z0063520 G1Z0244920 G1Z0048121 G1Z0083221 G1Z0094621 G1Z0094521 G1Z0115321 G1Z0123621 G1Z0190221 G1Z0190021 G1Z0190121 G1Z0190321 G1Z0210821 G1Z0210921 G1Z0211021
510(k) numbers
["K183291"]
PMA numbers
not on file
Product code
JAK
Firm FEI
1000513161
Firm city
Lexington
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register