RAYSEARCH LABORATORIES AB: RayStation 9.0.0.113 and 9.0.1.142. Radiation Therapy Treatment Planning System.

Recall number
Z-1744-2024
Recalling firm
RAYSEARCH LABORATORIES AB
Product
RayStation 9.0.0.113 and 9.0.1.142. Radiation Therapy Treatment Planning System.
Status
Open, Classified
Initiated
2024-03-28
Posted
2024-05-01
Terminated
not on file
Reason
Potential for reported SSD to be too high.
Root cause
Software design
Action
A Field Safety Notice for Medical Device Correction #133261, dated 3/28/24, was sent to consignees via email. Consignees were provided the instructions to use the Measure Tool to measure the distance in the patient view to make sure the reported SSD value is correct if it is to be used for patient setup or other critical tasks. If the SSD value is found to be incorrect, small changes to the isocenter, beam angles or the affected ROI's geometry can be made until the correct value is achieved. Consignees are to inform planning staff and all users of the workarounds provided in the Field Safety Notice. Consignees are to inspect their product and identify if installed units possess the affected software versions. Consignees are to reply to the recall notification confirming that they read and understood the Field Safety Notificaiton. RaySearch is to deploy a new version of RayStation/RayPlan to correct the issue, to be released in April 2024. The provided Field Safety Notice is to be maintained with affected units. Consignees with any questions are to contact quality@raysearchlabs.com.
Quantity
63 units
Distribution
Worldwide distribution - US Nationwide and the countries of Algeria, Australia, Austria, Bahrain, Belgium, Canada, China, Colombia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, France, Germany, Hungary, India, Indonesia, Iran, Israel, Italy, Japan, Korea, Malaysia, Morocco, Netherlands, New Zealand, Norway, Poland, Qatar, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UAE, UK.
Lot, serial or model codes
UDI: 0735000201017420190612 and 0735000201048820220420; GTIN: 07350002010174 and 07350002010488; Serial Numbers: 9.0.0.113 and 9.0.1.142; Expiration Date: 2027-12-15.
510(k) numbers
["K190387"]
PMA numbers
not on file
Product code
MUJ
Firm FEI
3007774465
Firm city
Stockholm
Firm state
not on file
Firm country
Sweden

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register