Ortho Development Corporation: UC Tibial Insert, Size 4, 13mm, REF 163-3413, Sterile R, Ortho Development, Draper, Utah 84020. The tibial insert is a device that fits into the tibial tray during knee revision. It provides the base for the metal femoral portion of the system, and moderates the degree of flexion afforded by the procedure.
- Recall number
- Z-1740-2011
- Recalling firm
- Ortho Development Corporation
- Product
- UC Tibial Insert, Size 4, 13mm, REF 163-3413, Sterile R, Ortho Development, Draper, Utah 84020. The tibial insert is a device that fits into the tibial tray during knee revision. It provides the base for the metal femoral portion of the system, and moderates the degree of flexion afforded by the procedure.
- Status
- Terminated
- Initiated
- 2011-01-07
- Posted
- not on file
- Terminated
- 2011-12-27
- Reason
- Components of a replacement knee system could be mislabeled.
- Root cause
- Error in labeling
- Action
- The firm, Ortho Development, sent a "Field Notification Memo" date January 7, 2011 to its customers. The memo described the product, problem and action to be taken. The customers were instructed to promptly return any remaining parts of the listed product; locate all the parts in their inventory; identify any of the listed parts that may have been used in surgery; verify the quantity they have on hand and the quantity used in surgery and reconcile them with the quantity sent; if discrepancy exists, notify customer service resolution; prepare items for return and contact customer service to obtain RMA number. Once reconciled, arrangements would be made for sending new inventory and the return of unused parts. If you have any questions regarding this issue, please notify Regulatory Affairs Manager at (801) 619-3450.
- Quantity
- 25 units
- Distribution
- Worldwide distribution: USA including states of: AZ, CA, FL, MI, TX, and UT; and country of: Japan.
- Lot, serial or model codes
- Lot Number: 670675.
- 510(k) numbers
- ["K060569"]
- PMA numbers
- not on file
- Product code
- JWH
- Firm FEI
- 3017298928
- Firm city
- Draper
- Firm state
- UT
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28