Beckman Coulter Inc.: Ultrasonic Transducer Kits For use with the Access Family of Immunoassay Systems*. *The Access Family of Immunoassay Systems includes the Access 2, UniCel DxI 800 and UniCel DxI 600, UniCel DxC 600i, and the UniCel DxC 880i, UniCel DxC 860i, UniCel DxC 680i, and UniCel DxC 660i systems.

Recall number
Z-1737-2014
Recalling firm
Beckman Coulter Inc.
Product
Ultrasonic Transducer Kits For use with the Access Family of Immunoassay Systems*. *The Access Family of Immunoassay Systems includes the Access 2, UniCel DxI 800 and UniCel DxI 600, UniCel DxC 600i, and the UniCel DxC 880i, UniCel DxC 860i, UniCel DxC 680i, and UniCel DxC 660i systems.
Status
Terminated
Initiated
2014-05-14
Posted
2014-06-09
Terminated
2014-12-25
Reason
Beckman Coulter is initiating a field action because some Ultrasonic Transducer Kits were potentially assembled with incorrect mounting screws.
Root cause
Mixed-up of materials/components
Action
Beckman Coulter sent an "URGENT MEDICAL DEVICE RECALL" letter dated May 8, 2014 via first class mail. The letter identified the reason for recall, product information, issue, impact, action, resolution, contact information, and customer response form. Customers are advised to review historical quality control (QC) data and System Check results, and determine whether a review of patient test results is warranted. For questions regarding this letter, please contact our Customer Support Center: From our website: http://www.beckmancoulter.com/customersupport/support, By phone: call 1-800-854-3633 in the United States and Canada, and Outside the United States and Canada, contact your local Beckman Coulter representative
Quantity
82 units
Distribution
Worldwide Distribution and USA Nationwide - and the countries of Australia, Canada, China, France, Hong Kong, Italy, Japan, Mexico, Portugal, Spain, Switzerland, Taiwan,and Turkey.
Lot, serial or model codes
Access 2 Immunoassay Systems, Catalog Number: 81600N. UniCel DxI 600 Access Immunoassay Systems, Catalog Numbers: A30260 and A71460. UniCel DxI 800 Access Immunoassay Systems, Catalog Numbers: 973100 and A71456. UniCel DxC 600i SYNCHRON Access Clinical Systems, Catalog Number: A25640.  Catalog numbers for these systems are included in the respective DxI 600 and DxI 800 model numbers: UniCel DxC  660i SYNCHRON Access Clinical Systems, UniCel DxC 680i SYNCHRON Access Clinical Systems, UniCel DxC 860i SYNCHRON Access Clinical Systems, UniCel DxC 880i SYNCHRON Access Clinical Systems.
510(k) numbers
["K042291"]
PMA numbers
not on file
Product code
JJE
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register