Skip to the content

FDA Medical Device Recalls (openFDA)

Integra LifeSciences Corporation: Integra PANTA Arthrodesis Nail System. Intended for use in tibiotalocalcaneal arthrodesis and treatment of trauma to the hindfoot and distal tibia.

Recall number
Z-1736-2013
Recalling firm
Integra LifeSciences Corporation
Product
Integra PANTA Arthrodesis Nail System. Intended for use in tibiotalocalcaneal arthrodesis and treatment of trauma to the hindfoot and distal tibia.
Status
Terminated
Initiated
2013-05-09
Posted
2013-07-16
Terminated
2013-11-07
Reason
As a result of the investigation into an adverse trend of complaints for reports of misalignment of the calcaneal or tibial screws during use of the Panta Nail Support Device, Integra has decided to issue this voluntary medical device correction. The misalignment can result in the inability to properly insert the screws into the Panta Nail and create the potential need for revision surgery.
Root cause
Device Design
Action
Integra initiated a voluntary correction on May 9, 2013 by providing a written notification and copy of the revised Surgical Technique Manual as well as the specific highlighted sections of the manual addressing calcaneal and tibial screw alignment to all consignees that Integra records indicate have been shipped a PANTA Nail. The written notifications and Surgical Technique Manual have been delivered to US Consignees by traceable courier service and by traceable email or facsimile for Canada and Non-US Consignees. The notification informs customers of the nature of the issue; advises of potential adverse patient consequences that may occur; advises customers to review highlighted revisions of the Surgical Technique Manual to minimize the potential for misalignment of the screws and requests customers to complete and return the attached Acknowledgement Return Form.
Quantity
692
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Austria, Australia, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Ireland, Israel, Italy, Netherlands, Portugal, Russia, Saudia Arabia, South Africa, Spain and United Arab Emirates.
Lot, serial or model codes
All lot numbers for PANTA Arthrodesis Nail Support Device Catalogue Nos: 519-110 and 519-130 from May 2006 to current.
510(k) numbers
["K050882","K091788"]
PMA numbers
not on file
Product code
HSB
Firm FEI
1121308
Firm city
Plainsboro
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register