Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer Biomet, Inc.: Lineum 4.35mm x 28mm Translation Screw, Model No. 14-524728 The Lineum OCT Spine System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion for acute and chronic instabilities of the craniocervical junction, also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.

Recall number
Z-1735-2019
Recalling firm
Zimmer Biomet, Inc.
Product
Lineum 4.35mm x 28mm Translation Screw, Model No. 14-524728 The Lineum OCT Spine System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion for acute and chronic instabilities of the craniocervical junction, also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
Status
Terminated
Initiated
2018-11-14
Posted
not on file
Terminated
2020-05-12
Reason
Etching on the tulip head incorrectly states "24MM" instead of the correct "28MM" marking.
Root cause
Process control
Action
On November 14, 2018, the firm sent an "Urgent Medical Device Recall" letters dated November 14, 2018, to its affected customers. The letter describes the product, problem and actions to be taken. The customers were instructed to do the following: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. Complete the included Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form must be returned even if you do not have affected products at your facility. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 5. If you have further questions or concerns after reviewing this notice, please call customer service at 1-800-447-3625 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com.
Quantity
64
Distribution
US Distribution to CA, NC, and NY; and International to Germany.
Lot, serial or model codes
Lot M26888
510(k) numbers
["K151224"]
PMA numbers
not on file
Product code
KWP
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register