FDA Medical Device Recalls (openFDA)
Angiodynamics, Inc.: ANGIODYNAMICS BENEPHIT(TM) XT Infusion System, 35 cm, Targeted Renal Therapy Infusion system, Model/REF Number 70035, STERILE, EO, Intended for Single Use Only. AngioDynamics, Incorporated 603 Queensbury, NY 12804.
- Recall number
- Z-1733-2010
- Recalling firm
- Angiodynamics, Inc.
- Product
- ANGIODYNAMICS BENEPHIT(TM) XT Infusion System, 35 cm, Targeted Renal Therapy Infusion system, Model/REF Number 70035, STERILE, EO, Intended for Single Use Only. AngioDynamics, Incorporated 603 Queensbury, NY 12804.
- Status
- Terminated
- Initiated
- 2010-04-30
- Posted
- 2010-05-28
- Terminated
- 2012-07-25
- Reason
- Benephit Infusion Catheters may exhibit a hole or tear in the primary sterile barrier packaging. The product is intended to be sterile. Breaches in the sterile packaging may result in non-sterile product, which could potentially cause a patient infection which is a serious medical condition.
- Root cause
- Packaging change control
- Action
- Angiodynamics, Inc. issued an "Urgent Medical Device Recall" notification (dated 4/28/10). Customers were advised of the affected product and instructed to return all recalled product to the firm. For further information, contact Angiodynamics, Inc. Customer Service at 1-800-772-6446.
- Quantity
- 741
- Distribution
- Distributed Nationwide and to Italy
- Lot, serial or model codes
- Batch Numbers: P909081, P909291, P910092, P910262, P911053, P911163, P911243, P912083, P912293 and P001203.
- 510(k) numbers
- ["K033569","K050205","K082163"]
- PMA numbers
- not on file
- Product code
- KRA
- Firm FEI
- 1319211
- Firm city
- Queensbury
- Firm state
- NY
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28