Skip to the content

FDA Medical Device Recalls (openFDA)

Angiodynamics, Inc.: ANGIODYNAMICS BENEPHIT(TM) XT Infusion System, 35 cm, Targeted Renal Therapy Infusion system, Model/REF Number 70035, STERILE, EO, Intended for Single Use Only. AngioDynamics, Incorporated 603 Queensbury, NY 12804.

Recall number
Z-1733-2010
Recalling firm
Angiodynamics, Inc.
Product
ANGIODYNAMICS BENEPHIT(TM) XT Infusion System, 35 cm, Targeted Renal Therapy Infusion system, Model/REF Number 70035, STERILE, EO, Intended for Single Use Only. AngioDynamics, Incorporated 603 Queensbury, NY 12804.
Status
Terminated
Initiated
2010-04-30
Posted
2010-05-28
Terminated
2012-07-25
Reason
Benephit Infusion Catheters may exhibit a hole or tear in the primary sterile barrier packaging. The product is intended to be sterile. Breaches in the sterile packaging may result in non-sterile product, which could potentially cause a patient infection which is a serious medical condition.
Root cause
Packaging change control
Action
Angiodynamics, Inc. issued an "Urgent Medical Device Recall" notification (dated 4/28/10). Customers were advised of the affected product and instructed to return all recalled product to the firm. For further information, contact Angiodynamics, Inc. Customer Service at 1-800-772-6446.
Quantity
741
Distribution
Distributed Nationwide and to Italy
Lot, serial or model codes
Batch Numbers: P909081, P909291, P910092, P910262, P911053, P911163, P911243, P912083, P912293 and P001203.
510(k) numbers
["K033569","K050205","K082163"]
PMA numbers
not on file
Product code
KRA
Firm FEI
1319211
Firm city
Queensbury
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register