Skip to the content

FDA Medical Device Recalls (openFDA)

International Technidyne Corp.: Hemochron Jr. ACT- LR Assay (JACT- LR); ITC Catalogue/Code Number: JACTLR 45 Cuvettes; For Performance on the HEMOCHROM Whole Blood Microagulation Systems; International Technidyne Corp., Edison, NJ 08820. The HEMOCHRON Jr. ACT-LR is a quantitative assay for monitoring heparin anticoagulation during various medical procedures. The ACT-LR demonstrated linear correlation to the anticoagulation effects of heparin up to 2.5 unit/ml of blood. It is intended for use in monitoring low to moderate heparin doses frequently associated with procedures such as cardiac catheterization, Extracorporeal Membrane Oxygenation (ECMO) Hemodialysis, and Percutaneous Transluminal Coronary Angioplasty.

Recall number
Z-1732-2009
Recalling firm
International Technidyne Corp.
Product
Hemochron Jr. ACT- LR Assay (JACT- LR); ITC Catalogue/Code Number: JACTLR 45 Cuvettes; For Performance on the HEMOCHROM Whole Blood Microagulation Systems; International Technidyne Corp., Edison, NJ 08820. The HEMOCHRON Jr. ACT-LR is a quantitative assay for monitoring heparin anticoagulation during various medical procedures. The ACT-LR demonstrated linear correlation to the anticoagulation effects of heparin up to 2.5 unit/ml of blood. It is intended for use in monitoring low to moderate heparin doses frequently associated with procedures such as cardiac catheterization, Extracorporeal Membrane Oxygenation (ECMO) Hemodialysis, and Percutaneous Transluminal Coronary Angioplasty.
Status
Terminated
Initiated
2009-05-05
Posted
2009-08-21
Terminated
2009-09-30
Reason
One lot of ACT-LR was released when one finished good testing criteria acceptance limit did not meet acceptance criteria.
Root cause
Employee error
Action
An "Urgent - Medical Device Recall" letter dated May 4, 2009 was sent to healthcare professionals. The recall letter described the reason for recall and the possible health risks involved. Consignees were asked to check their inventory for the affected lot number and return the completed Customer Account Tracking form via mail, e-mail or fax to ITC. ITC will make arrangements for the return of affected product. If there is no affected product in the inventory and no product or safety concerns, the letter is for information purposed only. Note: the Customer Account Tracking form should be completed in all cases. If you have questions regarding the recall information, please call ITC Technical Support at 1-800-579-2255, or email ITC at JACTLR@itcmed.com.
Quantity
173 boxes of 45 cuvettes
Distribution
Nationwide.
Lot, serial or model codes
Lot number K8JLR118.
510(k) numbers
["K960749"]
PMA numbers
not on file
Product code
JBP
Firm FEI
1000526865
Firm city
Edison
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register