Terumo Cardiovascular Systems Corporation: CDI Blood Parameter Monitoring System 500 with Arterial blood parameter modules and Hematocrit/Saturation probe. Intended for use during cardiopulmonary bypass procedures when continuous monitoring of arterial and/or venous pH, PCO2, PO2, K+, oxygen saturation, hematocrit, hemoglobin, and temperature are desired.

Recall number
Z-1731-2013
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
CDI Blood Parameter Monitoring System 500 with Arterial blood parameter modules and Hematocrit/Saturation probe. Intended for use during cardiopulmonary bypass procedures when continuous monitoring of arterial and/or venous pH, PCO2, PO2, K+, oxygen saturation, hematocrit, hemoglobin, and temperature are desired.
Status
Terminated
Initiated
2013-02-27
Posted
2013-07-13
Terminated
2016-02-17
Reason
Terumo discovered through its own testing that CDI 500 Blood Parameter Monitoring System software contains inaccuracies associated with a software algorithm, and a software update has been developed to address this issue and a premarket notification (K130425) has been submitted to the agency. Once clearance is received for this premarket notification from the US FDA, Terumo CVS will implement the
Root cause
Software design
Action
On 2/27/13, all consignees will receive a recall notification letter via Federal Express. Terumo CVS will correct all affected CDl Monitors. Correction to the devices will be completed concurrently with BPM replacements related to Z-2185/2189-11 RES 57194. Terumo continues to manufacture and ship devices affected by this recall. The long term corrective actions are associated with Recall# Z-2185/2189-11 RES 57194. As committed previously, we will re-notify all customers once we are ready to implement the fix for the BPM in the field. All units produced prior to the change in manufacturing will be included in the scope of this correction. Terumo continues to manufacturer units affected by this correction. All units returned for this correction will be tagged with a service traveler to indicate reason for correction and units will be properly segregated to prevent unintended re-shipment.
Quantity
3972 unit since 4/26/12
Distribution
Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, and WV and the countries of Jordan, CANADA, AUSTRALIA, Japan ,Hong Kong ,UNITED ARAB EMIRATES (UAE), BELGIUM, South Korea, Malaysia, Mexico, Philippines, Singapore, Taiwan, and Thailand.
Lot, serial or model codes
Catalog number: 500AHCT* and serial numbers: 1023,1025-1027,1029-1031, 1034,1035,1037,1040, 1045, 1048, 1054-1058, 1060-1062, 1064, 1065, 1072-1074, 1076-1095,1098-1100,1106-1110,1114,1116-1129,1131,1132,1141-1145, 1153-1165,1169-1180,1186-1191,1194-1210, 1212-1215,1221, 1230, 1237-1243,1249-1253,1256-1260,1265-1269,1275-1284,1290-1294,1300-1303, 1310, 1311, 1316-1320, 1322-1331,1335-1339,1345-1349,1359, 1378,1384-1388, 1394-1398,1409-1413,1419,1434-1438,1444-1448, 1454-1458, 1469-1478, 1481, 1482, 1484-1493,1495-1503,1505-1520,1528,1531-1553, 1555-1558,1567-1596,1607-1616,1632-1636,1647-1656,1667-1671,1677-1681,1687-1696, 1702-1711,1717-1721,1728-1732,1738-1753, 1759-1775,1777-1786,1790-1794,1796-1804,1810-1814,1820-1829,1831-1843,1847-1850, 1856, 1861-1885, 1891-1900, 1906-1915,1921-1925,1931-1935,1941-1950,1975,1986-2015, 2017-2022, 2024-2028, 2039, 2041-2044, 2046-2058, 2064, 2068, 2069, 2078-2085, 2087-2096, 2108, 2116, 2118- 2122, 2132-2137, 2143-2148, 2156-2176, 2183, 2184, 2188, 2191-2194, 2200-2213, 2215, 2220-2223, 2225-2227, 2232-2242, 2247, 2248, 2254, 2259,2261, 2262, 2264, 2269-2272, 2278, 2280-2284, 2292-2294, 2299-2312,2330-2337, 2353, 2359-2363, 2376, 2378-2388, 2398-2402, 2409-2416, 2427-2431, 2437, 2438, 2445-2452, 2459-2462, 2468, 2469, 2473-2475, 2484-2486, 2505-2513, 2518-2520, 2526-2540, 2557-2564, 2583-2586, 2604-2628, 2643-2645, 2648-2655, 2657-2660, 2663, 2684-2688,2692-2694,2697-2699, 2709, 2717-2721, 2728-2736, 2738-2741, 2748-2756, 2774-2782, 2785-2787, 2789-2792, 2796-2799, 2803, 2809-2814,2821-2836,2840-2845,2850-2859,2861-2864, 2871-2879, 2896-2900, 2903-2905, 2909-2912, 2916-2919, 2922, 2923, 2927-2936, 2941, 2948-2951, 2953-2957, 2976-2978, 2984-2989, 3000-3007, 3014-3021, 3028-3039, 3042-3051, 3056-3072, 3076-3082, 3088-3095, 3097-3114, 3117-3124,3126-3130,3136-3138,3147-3151, 3154,3163-3168,3183-3189, 3197-3213,3221-3224,3229-3231,3233-3237,3246,3247,3250-3260, 3269-3286,3290,3307-3313,3324-3328, 3335-3339, 3350-3353, 3358-3368, 3374-3382, 3392, 3393, 3398-3401, 3407-3409, 3411-3415, 3428-3433, 3438-3440, 3448, 3457, 3463-3468, 3473-3479,3483-3488,3491-3499,3504-3508, 3511-3517,3519-3533,3536-3540,3546-3570, 3574-3589,3610,3611,3620-3623,3632-3635, 3640-3642,3649-3660, 3662-3668,3682-3697, 3704-3714, 3717-3721,3727-3737,3745-3752, 3758-3768, 3781-3783, 3788-3798, 3805-3810, 3817-3826,3831-3837,3842-3851,3882-3891, 3900,3901,3907-3911,3922,3923,3932-3936, 3953-3961, 3966-3970, 3975-3985, 3990-3994, 3999-4005,4030-4037,4044-4049,4056-4069, 4076-4081, 4090-4095, 4100-4104, 4109-4113, 4133-4142,4149-4168,4186-4202,4213-4218, 4225-4229, 4242-4259,4281-4289,4300-4306, 4311-4317, 4324-4334, 4345-4350, 4353-4371, 5011-5035, 5040-5047, 5057-5064, 5078-5081, 5086-5093, 5102-5117, 5126-5141, 5150-5157, 5170-5177, 5180-5189, 5191-5196, 5199-5209, 5214-5221, 5226-5237, 5246-5249, 5254-5273, 5278-5281, 5286-5289, 5294-5297, 5306-5317, 5322-5329, 5345-5348, 5350-5353,5358-5369, 5375-5382, 5387, 5388, 5500-5515, 5527-5531, 5542-5556,5562-5564,5566-5577,5581-5583, 5593-5600, 5609-5660, 5662-5680, 5682-5688, 5690-5706, 5708-5713, 5715-5722, 5725-5727,and 5731.
510(k) numbers
["K972962"]
PMA numbers
not on file
Product code
DRY
Firm FEI
1828100
Firm city
Ann Arbor
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register