FDA Medical Device Recalls (openFDA)
Merit Medical Systems, Inc.: Vessel Sizing Angiographic Catheter, REF 7602-2M65, 5F PIG-VSC-2B, UHF Pigtail, Performa, Sterile EO, Merit Medical Systems, Inc., South Jordan, Utah 84095. Designed to be used for delivering radiopaque media to selected sites in the vascular system in conjunction with routine diagnostic procedures.
- Recall number
- Z-1731-2010
- Recalling firm
- Merit Medical Systems, Inc.
- Product
- Vessel Sizing Angiographic Catheter, REF 7602-2M65, 5F PIG-VSC-2B, UHF Pigtail, Performa, Sterile EO, Merit Medical Systems, Inc., South Jordan, Utah 84095. Designed to be used for delivering radiopaque media to selected sites in the vascular system in conjunction with routine diagnostic procedures.
- Status
- Terminated
- Initiated
- 2010-05-05
- Posted
- 2010-05-28
- Terminated
- 2010-06-01
- Reason
- Marker bands on angiographic catheters may not be properly attached, allowing movement of the bands.
- Root cause
- Process control
- Action
- Merit Medical Systems, Inc. began notifying sales representatives beginning May 5, 2010 via phone and instructed to contact all of their affected accounts. All affected product was to be returned to the firm. For further information, contact Merit Medical at 1-713-757-5067 or 1-801-208-4468.
- Quantity
- 96 units
- Distribution
- Nationwide Distribution -- GA, IL, MO, NY, PA and WI.
- Lot, serial or model codes
- Lot Number: G775032
- 510(k) numbers
- ["K000659"]
- PMA numbers
- not on file
- Product code
- DQO
- Firm FEI
- 1721504
- Firm city
- South Jordan
- Firm state
- UT
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28