FDA Medical Device Recalls (openFDA)
QIAGEN Gaithersburg, Inc.: Hc2 System Software Suite 4.0 version 3.0, available as component in Qiagen user guide, catalog #5050-00162, and as a component within a Qiagen PC, available as catalog # 5050-11140A. The software calculates and interprets assay results generated by the Digene Microplate Luminometer according to assay validation parameters. The software is used in conjunction with the following QIAGEN products: digene HC2 High-Risk HPV DNA Test [5199-1220 (one-plate); 5199-00016 (four-plate)] digene HC2 HPV DNA Test (5198-1220) digene HC2 CT/GC DNA Test (5130-1220) digene HC2 CT-ID DNA Test (5135-1220) digene HC2 GC-ID DNA Test (5140-1220) digene HC2 CT-GC Dual-ID DNA Test (5145-1220)
- Recall number
- Z-1728-2011
- Recalling firm
- QIAGEN Gaithersburg, Inc.
- Product
- Hc2 System Software Suite 4.0 version 3.0, available as component in Qiagen user guide, catalog #5050-00162, and as a component within a Qiagen PC, available as catalog # 5050-11140A. The software calculates and interprets assay results generated by the Digene Microplate Luminometer according to assay validation parameters. The software is used in conjunction with the following QIAGEN products: digene HC2 High-Risk HPV DNA Test [5199-1220 (one-plate); 5199-00016 (four-plate)] digene HC2 HPV DNA Test (5198-1220) digene HC2 CT/GC DNA Test (5130-1220) digene HC2 CT-ID DNA Test (5135-1220) digene HC2 GC-ID DNA Test (5140-1220) digene HC2 CT-GC Dual-ID DNA Test (5145-1220)
- Status
- Terminated
- Initiated
- 2011-01-12
- Posted
- not on file
- Terminated
- 2011-11-14
- Reason
- The software component of a diagnostic medical device may be defective and could cause incorrect patient values.
- Root cause
- Software design
- Action
- The firm, Qiagen, sent a "Notice of Product Field Action" letter dated January 12, 2011 to all customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately discontinue the process of running multiple assay protocols on a single microplate to eliminate the potential for inaccurate patient result reporting, and to complete and return the acknowledgment section via fax to QIAGEN 661-702-3854 or scan the document and email to techservice-na@qiagen.com. The firm informed users that they will provide further detailed instructions for a workaround and software fix. A QIAGEN representative will contact the customers who indicated that they run multiple assays on a plate and are currently using the HC2 System Software Version 3.0 (Suite 4.0), to assist in the identification of those patients that may have been affected by the software defect. If you have any questions or concerns, please contact Qiagen Technical Services at 800-344-3631, Option 2.
- Quantity
- 163
- Distribution
- The recalled product was sold to medical laboratories nationwide and in Canada.
- Lot, serial or model codes
- hc2 User's Guide, Lots 5723098 and 5722169; and Qiagen embedded hc2 Software Suite 4.0, Lots Q50455 and Q50349.
- 510(k) numbers
- ["K980120"]
- PMA numbers
- not on file
- Product code
- JJQ
- Firm FEI
- 3003572099
- Firm city
- Gaithersburg
- Firm state
- MD
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28