Skip to the content

FDA Medical Device Recalls (openFDA)

Exactech, Inc.: GXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. ACUMATCH 15 DEGREE LINER 22MM SZ C, REF 132-22-03 b. ACUMATCH 15 DEGREE LINER 28MM SZ D, REF 13-28-04

Recall number
Z-1726-2022
Recalling firm
Exactech, Inc.
Product
GXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. ACUMATCH 15 DEGREE LINER 22MM SZ C, REF 132-22-03 b. ACUMATCH 15 DEGREE LINER 28MM SZ D, REF 13-28-04
Status
Open, Classified
Initiated
2022-08-11
Posted
not on file
Terminated
not on file
Reason
Specific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the device and contribute to accelerated wear.
Root cause
Process control
Action
The firm sent an "URGENT MEDICAL DEVICE RECALL" DEAR HEALTHCARE PROFESSIONAL COMMUNICATION on 08/11/2022 by email. The purpose of the current letter is to update surgeons regarding the implications of the packaging non-conformity on performance of the Connexion GXL and other conventional UHMWPE acetabular liners. In addition, this letter provides updates to patient management and follow-up recommendations since issuance of the July 2021 DHCP letter. In addition to providing surgeons with lists of all their affected GXL and/or conventional polyethylene patients since 2004, Exactech is providing surgeons with two draft letters directed at patients who have been implanted with Exactech GXL hip devices. The Customers were instructed to take the following actions: CAREFULLY REVIEW THIS RECALL NOTIFICATION to ensure that you fully understand the issue identified, the recall strategy, and all actions required. IMMEDIATELY IDENTIFY AND QUARANTINE any of the subject devices in your inventory and/or customers inventory. EXTEND THE DESCRIPTION OF THE ISSUE AND CLINICAL IMPACT as described in the recall notification to your accounts that may have this product in their possession. COMPLETE AND RETURN the attached Recall Inventory Response Form to Exactech via email at recalls@exac.com. Reach out to Customer Service at cust.svc@exac.com to request a Returned Goods (RG) Number for ALL affected product requested to be returned. WE ARE REQUIRING 100% EFFECTIVENESS FOR THIS RECALL. If you have any questions, call (888) 912-0403.
Quantity
136 devices
Distribution
Worldwide distribution: US (nationwide) and OUS (foreign) countries of: ARGENTINA, BRAZIL, CHILE, CHINA, COLOMBIA, ECUADOR, GREECE, GUATEMALA, INDIA, LEBANON PORTUGAL,PUERTO RICO,SPAIN
Lot, serial or model codes
a. REF 132-22-03, GUIDID 10885862022882, Serial Numbers: 1191262, 1191263, 1191264, 1191265, 1191266, 1191267, 1191268, 1191269, 1191270, 1191271, 1191272, 1191273, 1264678, 1264679, 1264680, 1264681, 1264682, 1264683, 1264685, 1264686, 1264687, 1264688, 1264689, 1264690, 1264691, 1264692, 1264693, 1264694, 1264695, 1264697, 1264698, 1264699, 1264700, 1264701, 1264703, 1264704, 1632156, 1632157, 1632158, 1632159, 1632160, 1632161, 1644095, 1644096, 1644097, 1644098, 1644099, 1644100, 1644101, 1644102, 1644103, 1644104, 1644105, 1644107, 1644108, 1644109, 1644111, 1644112, 1644113, 1644114, 2668582, 2668587, 2668594, 2668606; b. REF 13-28-04, GUDID 10885862022998, Serial Numbers: 569977, 569980, 569983, 569985, 569986, 569987, 569988, 569992, 617606, 617607, 617608, 617609, 617611, 617615, 617616, 617619, 617620, 705092, 705094, 705099, 705100, 705101, 705103, 705114, 705117, 766031, 766032, 766034, 766036, 766038, 774540, 774543, 774547, 774548, 774550, 774552, 774553, 774554, 774555, 774556, 774558, 943503, 943504, 943505, 943506, 943519, 943521, 943522, 943525, 1067545, 1067546, 1067557, 1225656, 1225657, 1225658, 1225663, 1225664, 1225669, 1225679, 1225680, 1225683, 2781156, 4529972, 5247524, 5783760, 5783763, 5998768, 5998769, 5998770, 5998771, 6143949, 6143973.
510(k) numbers
["K040613"]
PMA numbers
not on file
Product code
MEH
Firm FEI
1038671
Firm city
Gainesville
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28