FDA Medical Device Recalls (openFDA)
HeartWare Inc: HeartWare Ventricular Assist System (HVAD). Catalog #: 1101, 1103, 1100, 1101, 1102, 1104, 1104JP, and 1205. Intended for use as a bridge to cardiac transplantation in patients who are at risk of death from refractory end-stage heart failure.
- Recall number
- Z-1726-2015
- Recalling firm
- HeartWare Inc
- Product
- HeartWare Ventricular Assist System (HVAD). Catalog #: 1101, 1103, 1100, 1101, 1102, 1104, 1104JP, and 1205. Intended for use as a bridge to cardiac transplantation in patients who are at risk of death from refractory end-stage heart failure.
- Status
- Open, Classified
- Initiated
- 2015-04-29
- Posted
- 2015-06-11
- Terminated
- not on file
- Reason
- Complaints with the HVAD Retraction of Pins within the driveline connector.
- Root cause
- Device Design
- Action
- Urgent Medical Device Correction letters, dated APR2015, were sent to users. The letters identified the affected device, as well as the reason for the recall and risks involved. Users were asked to review and familiarize themselves with the notice and "Patient Communication." The notice should be forwarded to anyone within customers' organization that need to be aware. Customers are to identify their patients who are currently supported by the HVAD System and distribute the "Patient Communication" to them. A HeartWare representative can assist with that process is needed. Customers should continue to reinforce the messages in the notice with their patients. The "Acknowledgement and Completion Form" should be completed, signed, and returned within 30 days of receipt of the notice. Questions or concerns should be directed to a local HeartWare representative. Clinical Support is also available at 1-888-494-6365.
- Quantity
- 3,747 currently in use
- Distribution
- Worldwide Distribution -- US, including the states of AL, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NH, NJ, NY, OH, OR, PA, SC, TN, TX, UT, and VA; and, the countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Israel, Italy, Japan, Korea, Lebanon, Lithuania, Luxembourg, Malaysia, Netherlands, New Zealand, Norway, Poland, Romania, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Turkey, and UK.
- Lot, serial or model codes
- Serial #: ALL HeartWare HVAD systems currently in use.
- 510(k) numbers
- not on file
- PMA numbers
- ["P100047"]
- Product code
- DSQ
- Firm FEI
- 3007042319
- Firm city
- Miami Lakes
- Firm state
- FL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28