Skip to the content

FDA Medical Device Recalls (openFDA)

Millar Instruments, Inc: Millar Instruments, Inc. Millar Sensors Systems Solutions Mikro-Tip Angiographic Catheters Sterile and Nonpyrogenic Model SPC-454D and SPC454F. Provide intravascular pressure readings and capability for the deliverance of high-speed injection of radiopaque contrast media used in vascular imaging to selected sites in the vascular system.

Recall number
Z-1725-2011
Recalling firm
Millar Instruments, Inc
Product
Millar Instruments, Inc. Millar Sensors Systems Solutions Mikro-Tip Angiographic Catheters Sterile and Nonpyrogenic Model SPC-454D and SPC454F. Provide intravascular pressure readings and capability for the deliverance of high-speed injection of radiopaque contrast media used in vascular imaging to selected sites in the vascular system.
Status
Terminated
Initiated
2011-03-08
Posted
2011-03-24
Terminated
2011-06-27
Reason
Possibility of epoxy debris particles inside the catheter lumen.
Root cause
Process design
Action
The firm, Millar, notified their consignees/customers by email and sent hardcopy of email on March 8, 2011. The notification described the product, problem and actions to be taken. The customers were instructed to return the catheter(s) to Millar or report back to Millar if the catheter(s) has been consumed. If you have any questions, please do not hesitate to contact Director, RA/QA at 832-667-7000.
Quantity
1080 units
Distribution
MN, NY, Washington DC, NC, MA, MN, PA, CA, MI, Australia, Germany, Canada, Japan, United Kingdom, and Norway.
Lot, serial or model codes
800-4219-1 and 800-4219-3 Listing #: D055900 and D055901
510(k) numbers
["K952773","K993694"]
PMA numbers
not on file
Product code
DQO
Firm FEI
1625382
Firm city
Houston
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register