FDA Medical Device Recalls (openFDA)
Elekta, Inc.: Synergy XVI Intended for radiation therapy treatment.
- Recall number
- Z-1724-2014
- Recalling firm
- Elekta, Inc.
- Product
- Synergy XVI Intended for radiation therapy treatment.
- Status
- Terminated
- Initiated
- 2014-04-23
- Posted
- 2014-06-16
- Terminated
- 2017-07-17
- Reason
- XVI can incorrectly calculate the target position of the treatment table.
- Root cause
- Software design
- Action
- Elekta is responsible for the correction of all products with this defect. 1. Notifications regarding the possibility of this software error and subsequent table location error and potential hazards have been sent out. The notifications include steps to take to reduce the possibility of the occurrence of an injury. 2. A software correction has been released for some versions of the XVI software. All versions of software with this error will be corrected by the end of July 2014. 3. The software correction will enable verification that values have been assigned to table position parameters; failure to pass verification will inhibit the system. The software correction will also detect internal communication errors (incomplete registrations) and inhibit the system when those errors are detected. 4. The update will be mandatory and will be made at no cost to the users. CDRH approves this CAP. You may proceed with implementation of the CAP. For further questions please call (770) 670-2422.
- Quantity
- 380 US
- Distribution
- Worldwide Distribution - USA including the states of AZ, CT, IA, OH, MD, MI, MN, NV, TX and VA., and the countries of India, Ireland, Malta, Netherlands, South Africa and UK.
- Lot, serial or model codes
- XVI R4.0 to R4.2.1, R4.5.0, R4.5.1, and R5.0.0 to R5.0.1
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- IYE
- Firm FEI
- 1037831
- Firm city
- Atlanta
- Firm state
- GA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as