Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Neuromodulation: Medtronic Model 8780 Ascenda Intrathecal Catheter. The implantable Medtronic Model 8780 Ascenda Intrathecal Catheter is part of an infusion system that stores and delivers parenteral drugs to the intrathecal space. The implanted infusion system components consist of a Medtronic pump and a Model 8780 Ascenda Intrathecal Catheter.

Recall number
Z-1723-2014
Recalling firm
Medtronic Neuromodulation
Product
Medtronic Model 8780 Ascenda Intrathecal Catheter. The implantable Medtronic Model 8780 Ascenda Intrathecal Catheter is part of an infusion system that stores and delivers parenteral drugs to the intrathecal space. The implanted infusion system components consist of a Medtronic pump and a Model 8780 Ascenda Intrathecal Catheter.
Status
Terminated
Initiated
2014-05-02
Posted
2014-06-04
Terminated
2014-10-07
Reason
Medtronic has determined that an Ascenda catheter kit may have had endotoxin in excess of USP limits.
Root cause
Release of Material/Component prior to receiving test results
Action
There was only one consignee affected by the recall. The consignee was contacted by phone on 5/2/14 and the information was provided verbally. General Manager & Vice President, Pain Stimulation/Targeted Drug Delivery spoke via phone with the physician managing the patient. The following information was provided: "Medtronic identified the potential for endotoxin levels exceeding USP requirements on product recently implanted (Model 8780, serial number N457062003). "There may be the potential for serious patient injury due to endotoxin within the first hours after implant. "Medtronic recommends observing the patient for symptoms of meningitis. Also to be aware that CSF would not test with a positive culture because endotoxin induced meningitis results in an aseptic meningitis. The physician expressed confidence in the management of the patient and stated he was calling the nursing care facility to increase monitoring of the patient. Subsequent information was received regarding the patient indicating there were no signs of an endotoxin reaction.
Quantity
1
Distribution
Distributed in the state of CT.
Lot, serial or model codes
N457062003
510(k) numbers
not on file
PMA numbers
["P860004S125"]
Product code
LKK
Firm FEI
2182207
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register