GE Medical Systems, LLC: Signa Openspeed 0.7T MR Systems, Model Numbers: Openspeed 3 (2138300-30), Openspeed 4 (2377062-5), Openspeed 5 (2377062-5, 2377062-30)

Recall number
Z-1723-2009
Recalling firm
GE Medical Systems, LLC
Product
Signa Openspeed 0.7T MR Systems, Model Numbers: Openspeed 3 (2138300-30), Openspeed 4 (2377062-5), Openspeed 5 (2377062-5, 2377062-30)
Status
Terminated
Initiated
2008-11-03
Posted
2009-08-25
Terminated
2012-05-24
Reason
GE Healthcare has identified a potential pinch point hazard on Ovation MR Scanner when using CTL Array XL Body Flex and Opened Body coil. A patient's hand may be pinched between the magnet bore ceiling cover and the CTL Array, XL Body Flex, or Open Body coil if the patient puts his/her hand on the top of the coil. Two injuries and four complaints filed. 2/20/09: Recall being expanded to incl
Root cause
Device Design
Action
GE Healthcare issued a "Product Safety Notification" letter dated October 23, 2008. The letter was addressed to Hospital Administrators, Mangers of Radiology/Cardiology & Radiologists/Cardiologists. The letter described the Safety Issue, Affected Product, Safety Instructions, Product Correction and Contact Information for the Signa Ovation 0.35T Systems. A second GE Healthcare "Urgent Medical Device Correction" letter dated March 5, 2009 was sent 3/9/09 to consignees which addressed the expanded recall to the Signa OpenSpeed0.7T MR Systems. For further questions, contact your local GE Healthcare sales or service representative.
Quantity
170(156 US / 14 OUS)
Distribution
Worldwide Distribution-- US (including states of AL, ALZ, AR, CA, CO,CT, DE, FL, FA, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MT, MO, NE,NH, NV, NJ, NY, NC, OH, OK, OR, PA, PR, SC SD, TN, TX, VA, WA, WV and WI), CHINA, CYPRUS, ARGENTINA, BRAZIL EGYPT, DENMARK, FEDERAL REPUBLIC OF GERMANY, FEDERAL REPUBLIC OF FRANCE, GET 7 D CHINA, GREECE, HUNGARY, ITALY, IVORY COAST, JAPAN, KOREA, KUWAIT, MALAWI, MEXICO, OMAN, PANAMA, PHILIPPINES, RUSSIAN FEDERATION, SAUDI ARABIA, SPAIN , SWITZERLAND, TURKEY, UKRAINE, UNITED ARAB EMERITUS, UNITED KINGDOM and VENEZUELA.
Lot, serial or model codes
Model Number: 2138300-30;  Serial Numbers: 00000203387MR6, 00000208208MR9, 00000216409MR3, 00000229302MR5, 00000229306MR6, 00000230309MR7, 00000231588MR5, 00000237930MR3, 00000239847MR7, 00000241694MR9, 00000247255MR3, 00000253103MR6, 00000189633MR1, 00000192104MR8, 00000193652MR5, 00000194218MR4, 00000194219MR2, 00000194950MR2, 00000196364MR4, 00000197334MR6, 00000198709MR8, 00000199443MR3, 00000201708MR5, 00000201709MR3, 00000201710MR1, 00000201901MR6, 00000202223MR4, 00000203386MR8, 00000204837MR9, 00000204838MR7, 00000204839MR5, 00000206438MR4, 00000206439MR2, 00000207493MR8, 00000207494MR6, 00000208209MR7, 00000208210MR5, 00000209712MR9, 00000209713MR7, 00000212511MR0, 00000212513MR6, 00000212514MR4, 00000212515MR1, 00000215366MR6, 00000216405MR1, 00000216406MR9, 00000216407MR7, 00000216409MR3, 00000217198MR1, 00000217201MR3, 00000221250MR4, 00000221251MR2, 00000221253MR8, 00000221254MR6, 00000222950MR8, 00000225273MR2, 00000225274MR0, 00000225942MR2, 00000226318MR4, 00000228022MR0, 00000228023MR8, 00000228024MR6, 00000228773MR8, 00000229303MR3, 00000229304MR1, 00000229305MR8, 00000230308MR9, 00000230312MR1, 00000231591MR9, 00000231592MR7, 00000232328MR5, 00000233644MR4, 00000233646MR9, 00000234832MR4, 00000234833MR2, 00000234834MR0, 00000237051MR8, 00000237052MR6, 00000237053MR4, 00000237926MR1, 00000237927MR9, 00000237928MR7, 00000237929MR5, 00000239480MR7, 00000239482MR3, 00000239483MR1, 00000239484MR9, 00000239845MR1, 00000239846MR9, 00000241126MR2, 00000241127MR0, 00000241693MR1, 00000241695MR6, 00000243405MR8, 00000244231MR7, 00000244232MR5, 00000247252MR0, 00000247253MR8, 00000247254MR6, 00000247256MR1, 00000250184MR9, 00000250185MR6, 00000250186MR4, 00000250187MR2, 00000253039MR2, 00000253102MR8, Unknown;  Model Number: 2377062-5;  Serial Numbers: 00000000100YR8, 00000000119YR8, 00000000129YR7, 00000000102YR4, 00000000103YR2, 00000000105YR7, 00000000106YR5, 00000000108YR1, 00000000109YR9, 00000000111YR5, 00000000113YR1, 00000000118YR0, 00000000120YR6, 00000000122YR2, 00000000124YR8, 00000000127YR1, 00000000128YR9, 00000000131YR3, 00000000132YR1, 00000000133YR9, 00000000134YR7, 00000000135YR4, 00000000136YR2, 00000000137YR0, 00000000142YR0, 00000000144YR6, 00000000146YR1, 00000000149YR5, 00000000150YR3, 00000000151YR1, 00000000152YR9, 00000000155YR2, 00000943610YM6, 00000951021YM5, 00000962592YM2;  Model Number: 2377062-30  Serial Numbers: 00000003047YR8, 00000003011YR4, 00000003022YR1, 00000003023YR9, 00000003024YR7, 00000003025YR4, 00000003026YR2, 00000003027YR0, 00000003028YR8, 00000003029YR6, 00000003030YR4, 00000003031YR2, 00000003033YR8, 00000003034YR6, 00000003035YR3, 00000003036YR1, 00000003039YR5, 00000003041YR1, 00000003042YR9, 00000003043YR7 and 00000003044YR5.
510(k) numbers
["K032795"]
PMA numbers
not on file
Product code
LNH
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register