FDA Medical Device Recalls (openFDA)

Medline Industries, LP: Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. LAPAROSCOPIC HERNIORRHAPHY DSC, Medline SKU # DYNJ63021B; 2. LAP PACK, Medline SKU # DYNJ64297B; 3. MINOR PROCEDURE PACK, Medline SKU # DYNJ82411.

Recall number
Z-1722-2026
Recalling firm
Medline Industries, LP
Product
Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. LAPAROSCOPIC HERNIORRHAPHY DSC, Medline SKU # DYNJ63021B; 2. LAP PACK, Medline SKU # DYNJ64297B; 3. MINOR PROCEDURE PACK, Medline SKU # DYNJ82411.
Status
Open, Classified
Initiated
2026-02-27
Posted
2026-04-07
Terminated
not on file
Reason
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Root cause
Under Investigation by firm
Action
Medline Industries, LP issued an "URGENT MEDICAL DEVICE RECALL" notice to its consignees on 2/27/2026 via USPS and email. The notice explained the issue, risk associated with syringe disconnection and loose connection, and request customers remove and destroy all Namic RA Syringes. Over-labels will be provided for any affected kits on hand stating that the affected syringes must be removed and discarded from further use. The only exception is in rare circumstances where angiography is urgently required, no alternative device is available, and postponing the procedure would place the patient at significant risk. Required Action: Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
Quantity
20 kits
Distribution
Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).
Lot, serial or model codes
Medline SKU # DYNJ63021B, UDI/DI each 10193489833775UDI/DI case 40193489833776, Lot Number: 26AMD887; Medline SKU # DYNJ64297B, UDI/DI each 10198459472992UDI/DI case 40198459472993, Lot Number: 26AMA970; Medline SKU # DYNJ82411, UDI/DI each 10195327193232UDI/DI case 40195327193233, Lot Number: 25IMB049.
510(k) numbers
not on file
PMA numbers
not on file
Product code
FDE
Firm FEI
1417592
Firm city
Northfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register