Accriva Diagnostics Inc., dba ITC, dba Accumetrics: Accriva- APTT Cuvette for use on the Hemochron Jr. test system, catalog #J103, lot B6JCA012.

Recall number
Z-1720-2016
Recalling firm
Accriva Diagnostics Inc., dba ITC, dba Accumetrics
Product
Accriva- APTT Cuvette for use on the Hemochron Jr. test system, catalog #J103, lot B6JCA012.
Status
Terminated
Initiated
2016-04-22
Posted
not on file
Terminated
2016-12-14
Reason
Clot activating reagent mis-positioned or visibly absent in the cuvette's test channel causing APTT test to either have an error code or an erroneously high result.
Root cause
Process design
Action
Accriva sent an " Urgent Field Safety Notice" letter dated April 22, 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were requested to take the following actions: 1. Forward communication to all those within organization who need to be aware of this matter. In addition, forward notification to all other organizations where affected devices may have been transferred to. 2. Check inventory to determine if you have any APTT Test Cuvettes, Lot Number: B6JCA012 3. If have inventory of APTT Test Cuvettes, with the subject Lot Number destroy the product and submit the attached Destruction Form along with your request for credit. 4. If have inventory DO NOT SHIP TO CUSTOMERS any kits labeled with APTT Test Cuvettes and the subject Lot Number. 5. If have shipped this product to customers please inform them immediately to STOP using the product. Provide the customers with a copy of this notification, request for the destruction of any unused product and the return of the attached forms. 6. Follow the instructions on the attached Field Corrective Action Form, complete the form and return it as instructed on the form. 7. If have questions, please contact Accriva Diagnostics directly contact information is provided below. Accriva Diagnostic, Attn: Regulatory Affairs USA Tel: 858.263.2347 Email: Fieldaction16-007@accriva.com
Quantity
135 boxes
Distribution
U.S. distribution to the following; state: VA. Foreign distribution to the following; China, Italy, Germany.
Lot, serial or model codes
Lot #B6JCA012
510(k) numbers
["K940401"]
PMA numbers
not on file
Product code
JPA
Firm FEI
2250033
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register