GE Medical Systems, LLC: GE Healthcare, CT Perfusion 4 on Advantage Workstation, CT and PET/CT consoles. Image analysis software package that allows the user to produce image data and to generate information with regard to changes in image intensity over time supporting CT Perfusion images.

Recall number
Z-1720-2009
Recalling firm
GE Medical Systems, LLC
Product
GE Healthcare, CT Perfusion 4 on Advantage Workstation, CT and PET/CT consoles. Image analysis software package that allows the user to produce image data and to generate information with regard to changes in image intensity over time supporting CT Perfusion images.
Status
Terminated
Initiated
2009-05-04
Posted
2009-08-13
Terminated
2012-05-16
Reason
GE Healthcare has recently become aware of the need to adjust default settings associated with the use of CT Perfusion 4 on the Advantage Workstation, ST and PET/CT consoles that may impact patient safety. The default parameter settings in CT Perfusion 4 may not be optimal in the case of data acquired with longer time sampling intervals such as axial scans with inter scan delay (ISD) greater th
Root cause
Software design
Action
GE Healthcare issued an "Urgent Medical Device Correction" letter dated April 23, 2009. The letter was addressed to Hospital Administrators / Risk managers and Radiology department mangers. The letter described the Safety Issue, Affected Product Details, Safety Instructions, Product Correction and Contact Information instructing the user to follow the enclosed User Guide Addendum to mitigate further problems. For further questions, contact GE Healthcare at 1-800-437-1171 (US), 0120-055-919 (Japan); for other countries, contact your local GE Healthcare field service representative.
Quantity
444
Distribution
Worldwide Distribution -- Including states of AZ, CA, CT, DC, FL, HI, IL, IA, MA, MI, NJ, NY, NC, OK, PA, PR, TX, WA, WV and WI and countries of ARGENTINA, BELGIUM, BOLIVIA, BRAZIL, CANADA, CHILE, CHINA, CZECH REPUBLIC, DENMARK, FINLAND, FRANCE, GERMANY, GREECE, GUATEMALA, HONDURAS, HONG KONG, INDIA, INDONESIA, ISRAEL, ITALY, JAPAN, REPUBLIC OF KOREA, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, PAKISTAN, POLAND, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN, TAIWAN, THAILAND, TURKEY, UNITED KINGDOM and VENEZUELA.
Lot, serial or model codes
Serial Numbers: 00000152741GE5, 00000152737GE3, 00000134590GE9, 00000119522GE1, 00000119523GE9, 00000146655GE6, 00000146668GE9, 00000146669GE7, 00000119519GE7, 00000145765GE4, 00000152738GE1, 00000121182GE0, 00000122793GE3, 00000119517GE1, 00000139086GE3, 00000141801GE1, 00000141802GE9, 00000141803GE7, 00000141804GE5, 00000141805GE2, 00000119521GE3, 00000118133GE8, 00000118134GE6, 00000118135GE3, 00000121180GE4, 00000141807GE8, 00000119516GE3, 00000123926GE8, 00000123931GE8, 00000122794GE1, 00000127524GE7, 00000127525GE4, 00000129247GE3, 00000134588GE3, 00000123927GE6, 00000123929GE2, 00000119520GE5, 00000122797GE4, 00000122795GE8, 00000123930GE0, 00000127522GE1, 00000127523GE9, 00000129243GE2, 00000127520GE5, 00000127521GE3, 00000127526GE2, 00000139088GE9, 00000129246GE5, 00000145645GE8, 00000141808GE6, 00000141811GE0, 00000151783GE8, 100995344, 00000145642GE5, 00000145647GE4, 00000118132GE0, 00000122799GE0, 00000122800GE6, 00000123928GE4, 00000127519GE7, 00000152735GE7, 00000129244GE0, 00000129245GE7, 00000129248GE1, 00000129249GE9, 00000129250GE7, 00000129251GE5, 00000134584GE2, 00000134585GE9, 00000134586GE7, 00000134587GE5, 00000145649GE0, 00000152742GE3, 00000122798GE2, 100995241, 100995217, 100995259, 80521349, 100995302, 100995275, 100995276, 100995431, 100995326, 100995412, 100995322, 80515302, 80515303, 80523248, 100995352, 100995298, 100995359, 100995306, 100995268, 100995354, 0CC08041501002, 0CC08041501003, 100995225, 80528332, 80521354, 100995279, 100995280, 100995281, 100995282, 100995283, 100995297, 100995305, 100995308, 100995288, 100995340, 100995355, 100995361, 100995385, 100995356, 100995450, 100995227, 80528736, 0CC08041501009, 80528724, 100995242, 100995351, 100995211, 100995212, 100995383, 80527216, 80527225, 80515730, 80528335, 80515721, 80521344, 80528730, 80528741, 80521352, 80521350, 80523251, 80521348, 100995222, 80515722, 100995339, 00000134592GE5, 00000134591GE7, 80505225, 80505229, 0CC08041501005, 80528328, 80528336, 80529188, 80521353, 100995224, 100995218, 100995221, 100995360, 100995240, 100995266, 100995267, 100995274, 80528329, 80528327, 00000121178GE8, 00000141809GE4, 00000152739GE9, 00000155218GE1, 00000121179GE6, 0CC08041501008, 100995239, 100995294, 100995327, 80528734, 100995366, 80515727, 80528322, 80521355, 100995278, 100995328, 100995331, 100995396, 100995411, 100995420, 00000146658GE0, 80528723, 80528324, 80505223, 80505226, 80505228, 100995220, 80528325, 0CC08041501007, 80527223, 80515296, 80528731, 80527219, 100995408, 0CC08041501010, 80528733, 80528735, 80528732, 80529189, 80521356, 100995311, 100995364, 100995219, 100995271, 100995292, 100995338, 100995334, 100995336, 100995286, 100995350, 100995379, 100995317, 100995318, 100995413, 100995449, 100995375, 80523250, 80515297, 80521351, 80528326, 0CC08041501004, 100995205, 80521343, 0CC08041501001, 100995270, 80505230, 80527218, 100995265, 100995300, 100995400, 100995426, 100995433, 80527224, 100995330, 100995341, 100995378, 100995291, 80528323, 0CC08041501006, 80505227, 100995377, 100995295, 100995272, 100995312, 100995368, 100995284, 100995349, 100995245, 80528330, 100995289, 100995255, 100995256, 100995374, 100995406, 100995319, 100995415, 100995409, 100995401, 100995332, 100995325, 100995419, 100995405, 100995407, 100995427, 100995429, 100995417, 80523253, 80528333, 00000145766GE2, 00000145644GE1, 80515298, 100995260, 0CC08040101001, 100995249, 100995388, 100995425, 100995229, 80523247, 100995209, 80528337, 100995296, 80515724, 80528339, 80505231, 80528739, 80515723, 80528340, 80515725, 80529195, 100995244, 80523249, 80515728, 80515729, 100995348, 100995370, 100995410, 100995376, 80529190, 80529191, 80529192, 80515726, 80529186, 80523246, 80528338, 80527222, 80529194, 80529193, 80528725, 80521346, 80528331, 80528334, 80528321, 80521347, 100995250, 100995251, 100995252, 100995254, 100995342, 000005189396-2, 100995223, 80523252, 100995215, 100995236, 100995231, 100995233, 100995232, 100995234, 100995237, 100995230, 100995235, 100995247, 100995304, 100995269, 100995273, 100995258, 100995285, 100995337, 100995287, 100995246, 100995314, 100995264, 100995387, 100995343, 100995365, 100995390, 100995392, 100995394, 100995391, 100995393, 100995384, 100995329, 100995404, 100995395, 100995402, 80528740, 80528737, 80527217, 80527221, 80528726, 80528727, 80528728, 100995316, 100995345, 100995353, 100995367, 80505224, 80505232, 100995207, 00000146661GE4, 00000144815GE8, 00000146660GE6, 00000146662GE2, 100995418, 100995441, 100995416, 00000146664GE8, 00000146666GE3, 00000146665GE5, 00000146659GE8, 100995238, 80529187, 100995208, 80523254, 100995214, 100995213, 100995206, 80515301, 80515305, 80527220, 100995253, 80523255, 100995210, 80515299, 80515304, 100995226, 100995257, 100995323, 100995228, 100995248, 100995262, 100995362, 100995301, 100995307, 100995309, 100995310, 100995299, 100995346, 100995403, 100995373, 100995363, 100995389, 100995422, 100995380, 100995381, 100995434, 00000145763GE9, 00000134589GE1, 00000121181GE2, 00000119518GE9, 00000151784GE6, 00000134583GE4, 00000141810GE2, 00000145641GE7, 00000145640GE9, 00000145768GE8, 00000145764GE7, 00000145767GE0, 00000151782GE0, 00000139087GE1, 00000145643GE3, 00000151780GE4, 00000151781GE2, 00000152740GE7, 00000152736GE5, 00000145646GE6, 00000145648GE2, 100995263, 100995243, 100995261, 100995421, 100995436, 100995435, 100995423, 100995424, 00000118131GE2, 00000141806GE0 and 00000118136GE1.
510(k) numbers
["K052839"]
PMA numbers
not on file
Product code
JAK
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register