Cook Incorporated: Roadrunner UniGlide Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): HPWS-35-150 ORDER NUMBER (GPN): G56152. For vascular catheter positioning and exchange in diagnostic and interventional procedures.

Recall number
Z-1717-2022
Recalling firm
Cook Incorporated
Product
Roadrunner UniGlide Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): HPWS-35-150 ORDER NUMBER (GPN): G56152. For vascular catheter positioning and exchange in diagnostic and interventional procedures.
Status
Open, Classified
Initiated
2022-07-19
Posted
not on file
Terminated
not on file
Reason
Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing, therefore, the sterility of the device may be compromised.
Root cause
Under Investigation by firm
Action
Cook notified consignees by Urgent Medical Device Recall Letter on July 19, 2022. Letter states reason for recall, health risk and action to take: Examine inventory immediately to determine if you have affected product(s) and quarantine any affected product that remains unused. Immediately cease all further distribution and use of the affected products. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Refer to the Acknowledgement and Receipt Form for return instructions. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter regardless of whether you have affected product(s) on hand. You may return the Form via fax (812.339.7316) or email (FieldActionsNA@CookMedical.com). 4. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected devices have been transferred. 5. Please immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: CustomerRelationsNA@CookMedical.com.
Quantity
30 units
Distribution
Nationwide Foreign: AR AT AU BE BG CA CA CH CL CN CZ DE DK ES FR GB GR GU HU IL IT JO JP KR KW MX MY NC NL PE PL PR PT PY RU SE SI TR UY ZA
Lot, serial or model codes
UDI-DI: (01)00827002561526- Lot Numbers: 14118842
510(k) numbers
["K110009"]
PMA numbers
not on file
Product code
DQX
Firm FEI
1820334
Firm city
Bloomington
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register