FDA Medical Device Recalls (openFDA)
Stryker Instruments Div. of Stryker Corporation: The NAV3i Platform Power Box supplies the electrical components with power (e.g. Stryker Nav3 Graphite Camera, SPC-3 Computer, IO-Tablet (touch screen) and the monitor).
- Recall number
- Z-1716-2014
- Recalling firm
- Stryker Instruments Div. of Stryker Corporation
- Product
- The NAV3i Platform Power Box supplies the electrical components with power (e.g. Stryker Nav3 Graphite Camera, SPC-3 Computer, IO-Tablet (touch screen) and the monitor).
- Status
- Terminated
- Initiated
- 2014-05-13
- Posted
- 2014-06-04
- Terminated
- 2014-09-15
- Reason
- The Stryker NAV3i Platform Power Boxes produced between December 9, 2013 and March 7, 2014 may have been assembled without electrical insulation of the soldering joints
- Root cause
- Process change control
- Action
- On May 13, 2014, Stryker Instruments sent URGENT MEDICAL DEVICE RECALL notification to their consignees. Notification included the description of the affected product, reason for the recall, risk to health, and actions to be taken by the Customer/User. Once Stryker receives the response from the customer, a Stryker Representative will contact the customer to set up a time to perform the upgrade to the affected power box. Old Power Boxes will be returned to Stryker and placed in an industrial compactor/recycler for destruction. Questions regarding the recall are directed to Kelly Jo Whipple M-F 8am-5pm @ 269-389-2921 or kellyjo.whipple@stryker.com
- Quantity
- 20
- Distribution
- Distributed in the states of AR, CA, IN, LA, MA, MD, MI, MT, NV, and TX.
- Lot, serial or model codes
- Stryker Product Number: 7700-853-000 Stryker Serial Numbers: 10004
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- HAW
- Firm FEI
- 1811755
- Firm city
- Portage
- Firm state
- MI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28