Olympus Corporation of the Americas: Brand Name: Powered Laser Surgical Instrument Product Name: SOLTIVE Pro SuperPulsed Laser System Model/Catalog Numbers: TFL-SLS, TFL-PLS Product Description: SOLTIVE Laser Systems use technology to increase procedural efficiency and flexibility to ureteroscopy, PCNL, and potential soft tissue applications. Component: No

Recall number
Z-1715-2025
Recalling firm
Olympus Corporation of the Americas
Product
Brand Name: Powered Laser Surgical Instrument Product Name: SOLTIVE Pro SuperPulsed Laser System Model/Catalog Numbers: TFL-SLS, TFL-PLS Product Description: SOLTIVE Laser Systems use technology to increase procedural efficiency and flexibility to ureteroscopy, PCNL, and potential soft tissue applications. Component: No
Status
Open, Classified
Initiated
2025-04-04
Posted
2025-05-01
Terminated
not on file
Reason
Preset treatment parameters are not consistently being used in accordance with the IFU.
Root cause
Unknown/Undetermined by firm
Action
On April 4, 2025 URGENT: MEDICAL DEVICE CORRECTION letters were sent to customers. Actions Required: Our records indicate that your facility has purchased one or more of the affected products. Olympus will contact you to schedule time for an Olympus Field Representative to visit your facility and install the software update. Additionally, Olympus requires you to take the following actions: 1. Carefully read the content of this notification. 2. Please check all areas of your facility to determine if you have the devices specified above. 3. Ensure all personnel are completely knowledgeable and thoroughly trained on the content of this notification and the SOLTIVE Laser System Instructions for Use. 4. Olympus requests that you acknowledge receipt of this letter through our recall web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number: "0469 c. Complete the form as instructed. 5. Olympus will contact you to schedule time for an Olympus Field Representative to visit your facility and install the software update. Olympus s record of the completed software update on your SOLTIVE system(s) will serve as the acknowledgment of this field corrective action for your facility. 6. Following the installation of the software update on your SOLTIVE system(s), ensure all personnel are thoroughly trained on the attached IFU Addendum corresponding with this update. The updated version of the full IFU can be located electronically at OlympusConnect.com. If you would like to receive a physical copy of the updated IFU, please submit a request in the comments of our recall portal when submitting your acknowledgement. 7. If you have further distributed this product, identify and forward them this notification. Olympus requests that you report any complaints related to the SOLTIVE Laser System or any associated injuries to our Technical Assistance Center (TAC) at 1- 800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may al
Quantity
189
Distribution
Worldwide distribution - US Nationwide and the countries of AU, CA, SG, DE, JP, HK.
Lot, serial or model codes
SOLTIVE Pro SuperPulsed Laser System: UDI-DI: TFL-SLS - 00821925044135; TFL-PLS - 00821925044111 All units included Powered Laser Surgical Instrument - Laser module component information Name: TFL Premium Laser Unit, TFL Standard Laser Unit UDI-DI: TFL-CPLU - 00821925044586; TFL-CSLU - 00821925044593
510(k) numbers
["K242191"]
PMA numbers
not on file
Product code
GEX
Firm FEI
2429304
Firm city
Center Valley
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register