Skip to the content

FDA Medical Device Recalls (openFDA)

Advanced Sterilization Products: EVOTECH Endoscope Cleaner and Reprocessing System 220-230V, P/N 50004-002. The ASP EVOTECH ECR Endoscope Cleaner and Reprocessor is a washer/disinfector, which is indicated for use with high-level disinfectant CIDEX OPA Concentrate and an enzymatic detergent (CIDEZYME Gl) to achieve cleaning and high level disinfection of heat sensitive (>60 C) semi-critical endoscopes.

Recall number
Z-1715-2015
Recalling firm
Advanced Sterilization Products
Product
EVOTECH Endoscope Cleaner and Reprocessing System 220-230V, P/N 50004-002. The ASP EVOTECH ECR Endoscope Cleaner and Reprocessor is a washer/disinfector, which is indicated for use with high-level disinfectant CIDEX OPA Concentrate and an enzymatic detergent (CIDEZYME Gl) to achieve cleaning and high level disinfection of heat sensitive (>60 C) semi-critical endoscopes.
Status
Terminated
Initiated
2015-05-11
Posted
2015-06-05
Terminated
2016-07-06
Reason
Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor (ECR) because it may not detect an improper endoscope connection on two channels, and subsequently pass the cycle without cancellation if there is a disconnected scope.
Root cause
Nonconforming Material/Component
Action
A customer notification letter dated 5/7/15 will be sent to all customers on 5/13/15 to inform them that ASP is recalling the EVOTECH Endoscope Cleaner and Reprocessor (ECR) because it may not detect an improper endoscope connection on two channels and subsequently pass the cycle without cancellation if there is a disconnected scope. The letter informs the customers of the problems identified and the actions to be taken. Customers are instructed to complete and return the attached Business Reply Card (double-sided) via mail or by fax to Stericycle: 888-879-8144. Customers with questions regarding the business reply card are instructed to contact Stericycle at 888-345-4671 and reference event #7781.
Quantity
44 units
Distribution
Worldwide Distribution-US (nationwide) and the countries of Australia, Singapore, Canada, and Thailand.
Lot, serial or model codes
5040207003 5040207004 5040308003 5042100008 5042100016 5042100017 5042100018 5042100019 5042100020 5042100021 5042100022 5042100023 5042100024 5042100025 5042100033 5042100034 5042100035 5042100036 5042100037 5042100038 5042100039 5042100040 5042100060 5042100061 5042100063 5042100064 5042100065 5042100066 5042100067 5042110002 5042110003 5042110004 5042110005 5042110006 5042120002 5042120006 5042130009 5042130011 5042140029 5042140030 5042140031 5042140032 5042140038 5042140039
510(k) numbers
["K140977"]
PMA numbers
not on file
Product code
FEB
Firm FEI
3003662624
Firm city
Irvine
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register