Stryker Instruments Div. of Stryker Corporation: The Stryker NAV3i Platform is a mobile cart platform consisting of the sub-components Stryker Nav3 Graphite Camera, SPC-3 Computer and IO-Tablet (touch screen) as well as the monitor, articulated arms and cart housing. This Navigation system is used in combination with surgical software for cranial, spine, ENT, orthopedic and trauma procedures.

Recall number
Z-1715-2014
Recalling firm
Stryker Instruments Div. of Stryker Corporation
Product
The Stryker NAV3i Platform is a mobile cart platform consisting of the sub-components Stryker Nav3 Graphite Camera, SPC-3 Computer and IO-Tablet (touch screen) as well as the monitor, articulated arms and cart housing. This Navigation system is used in combination with surgical software for cranial, spine, ENT, orthopedic and trauma procedures.
Status
Terminated
Initiated
2014-05-13
Posted
2014-06-04
Terminated
2014-09-15
Reason
The Stryker NAV3i Platform Power Boxes produced between December 9, 2013 and March 7, 2014 may have been assembled without electrical insulation of the soldering joints
Root cause
Process change control
Action
On May 13, 2014, Stryker Instruments sent URGENT MEDICAL DEVICE RECALL notification to their consignees. Notification included the description of the affected product, reason for the recall, risk to health, and actions to be taken by the Customer/User. Once Stryker receives the response from the customer, a Stryker Representative will contact the customer to set up a time to perform the upgrade to the affected power box. Old Power Boxes will be returned to Stryker and placed in an industrial compactor/recycler for destruction. Questions regarding the recall are directed to Kelly Jo Whipple M-F 8am-5pm @ 269-389-2921 or kellyjo.whipple@stryker.com
Quantity
20
Distribution
Distributed in the states of AR, CA, IN, LA, MA, MD, MI, MT, NV, and TX.
Lot, serial or model codes
Stryker Product Number: 7700-800-000 Stryker Serial Numbers: 15097, 15098, 15099, 15102, 15103, 15104, 15105, 15106, 15107, 15109, 15110, 15111, 15112, 15113, 15114, 15115, 15116, 15117, 15118
510(k) numbers
not on file
PMA numbers
not on file
Product code
HAW
Firm FEI
1811755
Firm city
Portage
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register