Skip to the content

FDA Medical Device Recalls (openFDA)

Greiner Bio-One North America, Inc.: VACUETTE Tube 2ml 9NC Coagulation Sodium Citrate 3.2%. Item no. 454322. Lot no. B1908355.

Recall number
Z-1714-2020
Recalling firm
Greiner Bio-One North America, Inc.
Product
VACUETTE Tube 2ml 9NC Coagulation Sodium Citrate 3.2%. Item no. 454322. Lot no. B1908355.
Status
Terminated
Initiated
2020-03-12
Posted
not on file
Terminated
2022-08-31
Reason
Concerned lot of VACETTE 2mL 9NC Coagulation Tubes were shown to have inconsistent amounts of the 2.3% sodium citrate.
Root cause
Process change control
Action
Greiner Bio-One North America Inc., notified customers on 03/12/2020 via "Urgent Product Recall" letter. The recall letter identified the affected products and requested the users to stop using the product and complete the product disposition site confirmation form. Consignees were asked to discard/destroy the affected products. If you have additional questions, please call our Technical Service at 800-815-8112.
Quantity
1,000,800 units.
Distribution
US nationwide distribution in the states of: NJ, IN, TN, MI, NC, WA, KS, CA, LA, IL, MD, NY, OH, TX, GA, KY, PA, CO, MA, PR, FL, AR, OK and NE.
Lot, serial or model codes
Item no. 454322. Lot no. B1908355. DI no. (01)29120017574488 (17) 200809 (10)B1908355.
510(k) numbers
["K971221"]
PMA numbers
not on file
Product code
GIM
Firm FEI
3001451379
Firm city
Monroe
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.