FDA Medical Device Recalls (openFDA)
Philips Healthcare: Philips Healthcare Brilliance iCT Computed Tomography X-Ray System
- Recall number
- Z-1714-2016
- Recalling firm
- Philips Healthcare
- Product
- Philips Healthcare Brilliance iCT Computed Tomography X-Ray System
- Status
- Terminated
- Initiated
- 2015-09-06
- Posted
- not on file
- Terminated
- 2021-03-04
- Reason
- Software Defects resulting in: (1) sagittal result shortened for axial scans; (2) single series displayed on console viewer when Creating multi-planner reconstruction; (3) scan length changes with a change in field of view; (4) DoseRight algorithm miscalculations leading to incorrect CTDI values; (5) scan length changes during subsequent axial results, (6) surview scan lengths near 135mm or 184mm
- Root cause
- Radiation Control for Health and Safety Act
- Action
- Philips plans to initiate a field corrective action that will install a new software release to address the issue. Philips will modify the systems in the field to correct the defect at no cost to the customer. 1. You will contact customers and initiate a software update to correct the defect. 2. You will install the software and perform testing to ensure the software update was effective. 3. The customer notification letter which includes a statement that you will without charge, remedy the defect or bring the product into compliance. For further questions, please call (1-800) 722-9377, Option 5
- Quantity
- US-423
- Distribution
- US Distribution - AZ, PA, FL, DC, W VA, NC, SC, AR, OK, OH, NY, NV, TX, PA, IN, CA, MN, MA, DC, MS, MD, OR, IL, CO, MO, VT, IA, AL, LA, GA and WA.
- Lot, serial or model codes
- Software v. 4.13, 4.14 and 4.15.
- 510(k) numbers
- ["K060937"]
- PMA numbers
- not on file
- Product code
- JAK
- Firm FEI
- 1218950
- Firm city
- Andover
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28