Skip to the content

FDA Medical Device Recalls (openFDA)

Arrow International Inc: Arrow Percutaneous Sheath Introducer Kits The Percutaneous sheath introducer products permit venous or arterial access and device introduction to the central circulation.

Recall number
Z-1713-2014
Recalling firm
Arrow International Inc
Product
Arrow Percutaneous Sheath Introducer Kits The Percutaneous sheath introducer products permit venous or arterial access and device introduction to the central circulation.
Status
Terminated
Initiated
2014-04-14
Posted
2014-06-04
Terminated
2016-10-14
Reason
Arrow is recalling the Arrow Percutaneous Sheath Introducer Kits due to dilator tip damage noted during internal inspection. Dilator tip damage has the potential to result in vessel damage.
Root cause
Process control
Action
Teleflex/Arrow International sent an Urgent Medical Device Recall Notification letter dated April 17, 2014, to distributors and end users. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately discontinue use and quarantine any affected product and to complete the enclosed Recall Acknowledgement Form and fax it to 1-800-343-2935 to return product. Customers would be contacted by a customer service representative for a Return Goods Authorization (RGA) Numbers and would be provided instructions for the return of the product. Customers should complete the form whether they have affected product or not. Customers with questions were instructed to call 1-800-343-2935. For questions regarding this recall call 610-378-0131.
Quantity
15424
Distribution
Worldwide Distribution - USA (nationwide) and Canada.
Lot, serial or model codes
Product codes and lot numbers: AK-09903-A 23F13G0162 AK-09903-S RF3064511 CDC-09903-1A 23F13G0642 AK-09903-J 23F13G0400 ASK-09903-CMC RF3064673 NL-09903-S 23F13G0630 AK-09903-LFSP RF3065449 ASK-09903-CMC 23F13G0343 NR-09903-S 23F13G0180 AK-09903-S 23F13G0204 ASK-09903-NKC 23F13G0336 SI-09903-E RF3064710 AK-09903-S RF3064956 ASK-09903-TJ 23F13G0567 SI-09903-E RF3064915 AK-09903-S 23F13H0195 CDC-09903-1A RF3065316
510(k) numbers
["K780532"]
PMA numbers
not on file
Product code
DYB
Firm FEI
3015859709
Firm city
Reading
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register