Beckman Coulter Inc.: Access 2 assay Immunoassay System, Remanufactured, Part No. 386220 Product Usage: The Access 2 and UniCel DxI Access Immunoassay Systems are in vitro diagnostic devices used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids.
- Recall number
- Z-1710-2017
- Recalling firm
- Beckman Coulter Inc.
- Product
- Access 2 assay Immunoassay System, Remanufactured, Part No. 386220 Product Usage: The Access 2 and UniCel DxI Access Immunoassay Systems are in vitro diagnostic devices used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids.
- Status
- Terminated
- Initiated
- 2017-03-07
- Posted
- 2017-03-15
- Terminated
- 2018-03-01
- Reason
- Beckman Coulter is recalling the Access 2 Family of Immunoassay Systems because a circuit board that controls motor functions can affect device positioning on the pipettor and incubator belt while the instrument is running. The position errors cause the instrument to stop running and cancel tests that are in process.
- Root cause
- Software Design Change
- Action
- An Urgent Medical Device Recall letter was sent on 3/7/17 to inform customers that Beckman Coulter has determined through internal investigation that a circuit board that controls motor functions in the instrument systems listed above can affect device positioning on the pipettor and incubator belt while the instrument is running. Customers are informed of the impact, action to be taken, and the resolution. Customers are instructed to complete and return the enclosed Response Form within 10 days. Customers with any questions are instructed to contact Customer Technical Support Center: From our website: http://www.beckmancoulter.com By phone: call 1-800-854-3633 in the United States and Canada. Outside the United States and Canada, contact your local Beckman Coulter representative.
- Quantity
- 851 units total for all products (291 units in US)
- Distribution
- Affected geographies: Algeria, Austria, Australia, Bangladesh, Bulgaria, Brazil, Canada, China, Costa Rica, Cyprus, Colombia, Czech Republic, El Salvador, France, Germany, Ghana, Greece, Guatemala, India, Indonesia, Italy, Japan, Jordan, Kazakhstan, Kenya, Lebanon, Lithuania, Malawi, Mexico, Morocco, Myanmar, Nepal, Oman, Panama, Philippines, Pakistan, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, Saudi Arabia, Serbia, South Africa, South Korea, Spain, Switzerland, Tunisia, Turkey, United Kingdom, United States, Vietnam, Zimbabwe
- Lot, serial or model codes
- Serial No. 501012,501110,501190,501224,501294,501371,501394,501403,501440, 501443,501490,501501,501673,501681,501693,501861,501920,501926, 501936,501970,502165,502196,502319,502331,502332,502368,502458, 502475,502506,502543,502549,502556,502574,502640,502647,502651, 502791,502843,502848,503153,503279,503313,503354,503384,503392, 503424,503543,503627,503678,504046,504071,504072,504092,504112, 504195,504322,504365,504370,504756,504767,504902,505042,505195, 505251,505254,505644,505795,505840,505947,506002,506026,506108, 506119,506153,506187,506371,506406,506560,506586,506664,506708, 506766,506786,506837,506846,507039,507120,507161,507162,507214, 507229,507248,507282,507353,507397,507480,507511,507604,507619, 507842,507929,508019,508096,508179,508227,508313,508334,508412, 508478,508509,508516,508767,509051,509087,509113,509164,509253, 509297,509632,509837,510049,511042,511047,511180,511383,700024, 800321,800341,800443,800608,800611,800650,800849
- 510(k) numbers
- ["K922823"]
- PMA numbers
- not on file
- Product code
- JJE
- Firm FEI
- 2050012
- Firm city
- Brea
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28