Skip to the content

FDA Medical Device Recalls (openFDA)

Nidek Inc: EC-5000 Excimer Laser Corneal Surgery System (EC-5000, EC 5000 CXIII, EC-5000 Quest. Opthalmic: NIDEK EXCIMER LASER CORNEAL SURGERY SYSTEM, model EC-5000, is an ophthalmic laser system used for correction of corneal refraction and ablation of the corneal surface.

Recall number
Z-1710-2015
Recalling firm
Nidek Inc
Product
EC-5000 Excimer Laser Corneal Surgery System (EC-5000, EC 5000 CXIII, EC-5000 Quest. Opthalmic: NIDEK EXCIMER LASER CORNEAL SURGERY SYSTEM, model EC-5000, is an ophthalmic laser system used for correction of corneal refraction and ablation of the corneal surface.
Status
Terminated
Initiated
2015-03-24
Posted
2015-06-04
Terminated
2016-04-05
Reason
Multi-stage treatment option for the EC-5000 Operator's Manual was not approved by the FDA.
Root cause
No Marketing Application
Action
Nidek sent an Device Correction letter dated April 30th, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Nidek will contact customers to schedule an appointment to replace the operators manual and implementing a software lockout device on the device so that the multi stage feature is no longer available. If a Nidek rep does not contact you before April 30, 2015, please call 800-223-9044 to schedule an appointment, All non responding customers.will be contacted three time via Fed Ex letter
Quantity
114 units
Distribution
Us distribution only.
Lot, serial or model codes
Alll devices currently in use by consignees.
510(k) numbers
not on file
PMA numbers
["P970053"]
Product code
LZS
Firm FEI
2936921
Firm city
Fremont
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register