Skip to the content

FDA Medical Device Recalls (openFDA)

Medline Industries, LP: Medline kits containing Olympus biopsy valves: 1. BRONCHOSCOPY, DYNJ900898I DYNJ901922G 2. FLEXIBLE BRONCHOSCOPY KIT, DYKM1417I DYKM1417J 3. LARYNGOSCOPY/BRONCHOSCOPY PACK, DYNJ58146A 4. RIGID KIT, DYNDA2138A

Recall number
Z-1709-2026
Recalling firm
Medline Industries, LP
Product
Medline kits containing Olympus biopsy valves: 1. BRONCHOSCOPY, DYNJ900898I DYNJ901922G 2. FLEXIBLE BRONCHOSCOPY KIT, DYKM1417I DYKM1417J 3. LARYNGOSCOPY/BRONCHOSCOPY PACK, DYNJ58146A 4. RIGID KIT, DYNDA2138A
Status
Open, Classified
Initiated
2026-02-16
Posted
2026-04-01
Terminated
not on file
Reason
Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
Root cause
Under Investigation by firm
Action
Firm notified customers on 2/16/26. Consignees were provided with a copy of Olympus's notification, which includes reminders to follow the IFU. According to the firm, the component can continue to be utilized by the customer, along with the remainder of the kit.
Quantity
14,379 kits total
Distribution
US Nationwide distribution in the states of CA, FL, IL, IN, KY, MA, MD, MT, NY, OH, PA, TN, TX.
Lot, serial or model codes
1. DYNJ900898I UDI-DI 10198459273513 (EA) 40198459273514 (CS) LOTS 25DME064 25HMD841 25JMD489 25KME329 25KMH848 25LMC029 25LMH235 25LMK305 DYNJ901922G UDI-DI 10195327640422 (EA) 40195327640423 (CS) LOTS 24KBP812 2. DYKM1417I UDI-DI 10195327456757 (EA) 40195327456758 (CS) LOTS 24DLA405 24ELA912 24GLA871 24ILA151 24JLA326 DYKM1417J UDI-DI 10198459191640 (EA) 40198459191641 (CS) LOTS 25BLA761 25BLA886 25DLA353 25ELA056 25FLA429 25HLA196 25HLA242 25KLA388 25KLA610 3. DYNJ58146A UDI-DI 10193489611977 (EA) 40193489611978 (CS) LOTS 23KDA851 24ADC096 24CDB107 24DDB887 24IDA855 24JDB494 24LDA691 25BMF825 25DMD969 25FMC760 25GMD422 4. DYNDA2138A UDI-DI 10193489874594 (EA) 40193489874595 (CS) 24DBM325 24DBM383 25DBK159
510(k) numbers
not on file
PMA numbers
not on file
Product code
OGR
Firm FEI
1417592
Firm city
Northfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register