Stryker Sustainability Solutions: Reprocessed Daig Supreme Fixed Curve Diagnostic Electrophysiology Catheters; Model #: 401450, 401877. Indicated for temporary intracardiac sensing, recording, stimulation, and electrophysiology mapping of cardiac structures.

Recall number
Z-1708-2016
Recalling firm
Stryker Sustainability Solutions
Product
Reprocessed Daig Supreme Fixed Curve Diagnostic Electrophysiology Catheters; Model #: 401450, 401877. Indicated for temporary intracardiac sensing, recording, stimulation, and electrophysiology mapping of cardiac structures.
Status
Terminated
Initiated
2016-04-28
Posted
not on file
Terminated
2016-06-29
Reason
Mislabeled for size
Root cause
Under Investigation by firm
Action
The firm sent out customer notification letters on 04/28/16. Customers are instructed to discontinue the use of affected lots of recalled product. Other EP Catheters reprocessed by Stryker Sustainability Solutions should be considered as alternatives. A Recall Effectiveness Check Form including the quantity of affected devices shipped to facility is enclosed. Please complete the Recall Effectiveness Check Form and indicate if any devices from the affected lots remain in your inventory. This form needs to be completed even if no affected product is found. Please return the completed and signed Recall Effectiveness Check Form to your local Stryker Sustainability Sales Representative or email to SSSPFA@stryker.com or mail to: Stryker Sustainability Solutions 1810 West Drake Drive Tempe, AZ, 85283 Attn: Jodie Rueckert If the form indicates that affected devices remain in inventory, a prepaid shipping label will be issued for the return of the product. Customers will receive a credit for all affected devices returned. Adverse reactions or quality problems experienced with the use of this product may be reported to: - Stryker Sustainability Solutions Complaint Hotline: +1(888) 888-3433 X5555 - http://www.stryker.com/productexperience/ - The FDAs MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. On 05/12/16 the firm sent out a clarification letter to emphasize that use of affected lots of product should be discontinued. In addition, SSS would like to note that, use of devices larger than expected may result in excess force applied to the blood vessel. The risk to the patient can vary from a prolongation in procedure time, vessel rupture, organ damage or other unforeseen consequences. Some of these injuries may require surgery or other procedures to remedy. Any questions contact the Stryker Sustainability Solutions Sales Representative or Regulatory Affairs.
Quantity
18 devices
Distribution
Distributed in the state of NY and the country of Canada.
Lot, serial or model codes
Catalog Number: 401450 Lot Number: 2561683E Expiration Date: 5/13/2017  Catalog Number: 401877 Lot Number: 2555866E Expiration Date: 5/6/2017
510(k) numbers
["K012708"]
PMA numbers
not on file
Product code
NLH
Firm FEI
3009350690
Firm city
Tempe
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register