Philips Electronics North America Corporation: Philips IntelliVue MX4O WLAN Patient Wearable Monitor Product: 865352 Exchange part (service numbers): 453564615311 TELE PWM,802.lla/b/g,ECG only, US only 453564615331 TELE PWM,802.1 1 a/b/g,ECG&Sp02, US only Product Usage: Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults and pediatrics in hospital environments and during transport inside hospitals. The MX4O is to be used primarily as a traditional telemetry medical device. It connects to the IntelliVue Information Center by way of a wireless network. When the MX4O is connected the IntelliVue Information Center the IntelliVue Information Center provides the primary patient monitoring and alarming function. The MX4O does not automatically provide local monitoring or alarming when connected to the Information Center.

Recall number
Z-1707-2017
Recalling firm
Philips Electronics North America Corporation
Product
Philips IntelliVue MX4O WLAN Patient Wearable Monitor Product: 865352 Exchange part (service numbers): 453564615311 TELE PWM,802.lla/b/g,ECG only, US only 453564615331 TELE PWM,802.1 1 a/b/g,ECG&Sp02, US only Product Usage: Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults and pediatrics in hospital environments and during transport inside hospitals. The MX4O is to be used primarily as a traditional telemetry medical device. It connects to the IntelliVue Information Center by way of a wireless network. When the MX4O is connected the IntelliVue Information Center the IntelliVue Information Center provides the primary patient monitoring and alarming function. The MX4O does not automatically provide local monitoring or alarming when connected to the Information Center.
Status
Terminated
Initiated
2017-03-06
Posted
not on file
Terminated
2018-05-25
Reason
Philips lntelliVue MX4O WLAN Patient Wearable Monitor may not automatically switch to Monitor Mode with audible alarms when association with central monitoring system is unsuccessful [incomplete]
Root cause
Software design
Action
Philips issued Field Safety Notice on 3/6/17. The Field Safety Notice informs customers of the issue, identifies details of the units affected, gives instructions on actions to be taken by the customer .Philips is asking customers to follow the Action to be Taken by Customer/User section of the Field Safety Notice: When a No Data Tele INOP alert is displayed on a Philips IntelliVue Information Center iX (PIIC iX), check the MX40 as directed in the PIIC iX B.01 IFU, page 87. If the MX40 is in the Coverage Area, remove the battery, and reinsert the battery. If the MX40 is not in the Coverage Area, place the sector in the PIIC iX in standby as directed in the MX40 B.06 IFU, page 53, or return the patient and MX40 to the Coverage Area. Philips is releasing a software upgrade B.06.18 for affected Philips IntelliVue MX40 Patient Wearable Monitors. A Philips Healthcare representative will contact customers with affected devices to arrange for the installation of the software upgrade
Quantity
2648
Distribution
Worldwide Distribution - US Nationwide including the countries of: Australia Belgium Canada Denmark Finland France Germany Germany India Italy Netherlands Norway Qatar Singapore South Africa Sweden Switzerland United Kingdom
Lot, serial or model codes
The following product number and Exchange part numbers containing the combination of HW Revision C.01 .01 with SW Revisions B.05 or B.06 of the Philips IntelliVue MX4O Patient Wearable Monitor
510(k) numbers
["K113125"]
PMA numbers
not on file
Product code
DSI
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register