FDA Medical Device Recalls (openFDA)
Angiodynamics, Inc.: PULSESPRAY INFUSION SYSTEM, 90 cm x 20 cm infusion length, Catalog No./Ref 12500402, STERILE, EO. AngioDynamics, 603 Queensbury Avenue, Queensbury, NY 12804. Infusion Catheter.
- Recall number
- Z-1707-2010
- Recalling firm
- Angiodynamics, Inc.
- Product
- PULSESPRAY INFUSION SYSTEM, 90 cm x 20 cm infusion length, Catalog No./Ref 12500402, STERILE, EO. AngioDynamics, 603 Queensbury Avenue, Queensbury, NY 12804. Infusion Catheter.
- Status
- Terminated
- Initiated
- 2010-05-17
- Posted
- 2010-05-27
- Terminated
- 2012-07-25
- Reason
- This lot was packaged with the wrong occluding ball guidewire. The PulseSpray Catheter is 90 cm in length and requires a 109 cm occluding ball guidewire, however this lot was packaged with a 60 cm occluding ball guidewire.
- Root cause
- Process control
- Action
- Angiodynamics issued an "Urgent Medical Device Recall" notification (dated 5/11/10) to user accounts via Certified Mail. Consignees were asked to identify, segregate and return all affected product to the firm in addition to sending in a completed Reply Form via fax. For further information, contact Angiodynamics Customer Service at 1-800-772-6446.
- Quantity
- 20 units
- Distribution
- Worldwide Distribution -- United States (AR, NY, PA, MI, MN, LA and WA), Spain, Austria and Denmark.
- Lot, serial or model codes
- Lot Number: 991437.
- 510(k) numbers
- ["K905447"]
- PMA numbers
- not on file
- Product code
- KRA
- Firm FEI
- 1319211
- Firm city
- Queensbury
- Firm state
- NY
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28