Skip to the content

FDA Medical Device Recalls (openFDA)

SUMMA THERAPEUTICS, LLC: Brand Name: Finesse BTK Multicath Product Name: Injectable Angioplasty Balloon Catheter Model/Catalog Number: FS252251502 Product Description: 5 French sheath, 0.014" guidewire compatible over-the-wire injectable angioplasty balloon catheter, balloon diameter 2.5mm, balloon length 225mm, catheter working length 150cm Component: NA

Recall number
Z-1704-2026
Recalling firm
SUMMA THERAPEUTICS, LLC
Product
Brand Name: Finesse BTK Multicath Product Name: Injectable Angioplasty Balloon Catheter Model/Catalog Number: FS252251502 Product Description: 5 French sheath, 0.014" guidewire compatible over-the-wire injectable angioplasty balloon catheter, balloon diameter 2.5mm, balloon length 225mm, catheter working length 150cm Component: NA
Status
Open, Classified
Initiated
2025-07-21
Posted
2026-03-30
Terminated
not on file
Reason
Potential for the balloon in the device to not meet burst specifications.
Root cause
Process control
Action
On July 21, 2025, customers were notified via E-mail. We discovered that the burst test values recorded did not meet the specification of the safety rated burst test average of 17atm. Please review your inventory and we ask that these catheters be returned for further investigation.
Quantity
22 units
Distribution
US Nationwide distribution in the states of New Jersey, Florida.
Lot, serial or model codes
Model/Catalog Number: FS252251502; UDI-DI: 00810017490967; lots: 240502, 240095, 231296, 231296A and 240095A;
510(k) numbers
["K230263","K241740"]
PMA numbers
not on file
Product code
LIT
Firm FEI
3027395815
Firm city
Allston
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register